background image

ENGLISH - 2/8

 

 TECHNICAL SPECIFICATIONS 

STANDARDS

 

EN61000-6-4

 electromagnetic emissions, industrial environment  

EN61000-6-2

. Electromagnetic immunity, industrial environment.  

EN61010-1

 (safety)

 

Install a fuse with a maximum capacity of 2.5 A near the module  

For voltages over 1000 V , install a 4kV overvoltage protection device 

INSULATION

 

Protection 

impedance

 

N.B: use in environments with pollution degree 2 or lower.

 

ENVIRONMENTAL 

CONDITIONS

 

Temperature: -20 – + 65°C

 

Humidity/Altitude: 30 -90% non condensing, up to 2000 m above sea level.  

Storage temperature: -20 – + 85°C

 

Protection rating: IP20

 

CONNECTIONS

 

Removable 4-way screw terminals, 3.5 mm pitch for cable up to 2.5 

mm2, rear IDC10, for 35 mm DIN bar IEC EN60715 

3.5 mm front jack for configuration  

4 mm standard bushes

 

POWER SUPPLY

 

Voltage 10..40 V

 

or 19..28 V

50 – 60 Hz

 

Typical absorption 1 W

 

INPUTS

 

Voltage: 0/850 V

~

 

Input impedance 4 MΩ (precision: 0.5%). 

 

Frequency 30 - 60 Hz,

 

Voltage: 0/1200 V , 

Input impedance 4 MΩ (precision:

 0.5%).  

Pass-band: 30-400 Hz (-3 dB).

 

N.B.:

 the instrument below 3% of the selected full scale does not give the 

correct measurement

OVERVOLTAGE

 

Category V: up to 300 Vac/dc:

 

Category III: up to 600 Vac/dc;  

Category II: up to 1000 Vac/dc;

 

For voltages up to 1200 V  provide an overvoltage limiter external to the 4 

kV device. 

OUTPUTS

 

Current: 0/4..20 mA, max resistance 500 Ω (precision: 0.1%)

 

Continuous voltage: 0-10 V selectable via the software. Minimum load 

resistance: 1 kΩ. (precision: 0.1%)

 

Temperature drift: 100 ppm/K

 

Response time for a stepped variation: 1 s from 10 to 90%

 

CONFIGURABILITY

 

Baud rate and MODBUS Address via DIP-switch or via software.

 

Dimensions (LxHxD)

 

35 x 102.5 x 111 mm (terminals included).

 

Case

 

PA6 material, black.

 

t

Power

Supply

4000 V

~

1500 V

~

1

0.

.40 Vdc / 19..28 Vac

0..1200 Vdc

0..850   Vad

Input

Configuration

port

Output

RS232

RS485

mA / V

Содержание Z204-1

Страница 1: ...tated data may be modified or supplemented for technical and or sales purposes INSTALLATION MANUAL Z204 1 True RMS AC DC converter with Modbus protocol on RS485 SENECA s r l Via Austria 26 35127 PADOVA ITALY Tel 39 049 8705355 8705359 Fax 39 049 8706287 For manuals in French German English and configuration software visit the www seneca it products 204 1 website EN THE ORIGINAL VERSION IS IN ITALI...

Страница 2: ... 28 V 50 60 Hz Typical absorption 1 W INPUTS Voltage 0 850 V Input impedance 4 MΩ precision 0 5 Frequency 30 60 Hz Voltage 0 1200 V Input impedance 4 MΩ precision 0 5 Pass band 30 400 Hz 3 dB N B the instrument below 3 of the selected full scale does not give the correct measurement OVERVOLTAGE Category V up to 300 Vac dc Category III up to 600 Vac dc Category II up to 1000 Vac dc For voltages up ...

Страница 3: ... www seneca it prodotti Z204 1 The module must be repaired and damaged parts replaced by the Manufacturer The product is sensitive to electrostatic discharges Take appropriate measures during any operation Important Obstructing ventilation slots with any object is prohibited Installing the module next to devices that generate heat is prohibited Electrical and electronic waste disposal applicable i...

Страница 4: ... 14 15 16 1 With the help of a screwdriver pull the two hooks on the side of the module outwards 17 18 19 20 25 26 27 28 21 22 23 24 29 30 31 32 2 Slowly extract the module from the rail USE OF THE Z PC DINAL ACCESSORY CAUTION Do not turn the module upside down and do not force the insertion of the IDC10 connector on the Z PC DIN bus The module s rear IDC10 connector must be plugged into a free sl...

Страница 5: ...RICAL CONNECTIONS WARNING BEFORE CARRYING OUT ANY INSTRUMENT CONNECTIONS MAKE SURE THAT YOU DISCONNECT ALL THE CIRCUITS SUBJECT TO A DANGEROUS VOLTAGE TO CONNECT THE HIGH VOLTAGE INPUT USE ONLY THE PLUGS SUPPLIED WITH THE INSTRUMENT Insertion of high voltage plugs The figure at the side shows the insertion points of the two 4mm banana plugs supplied with the instrument Polarity of the high voltage...

Страница 6: ...ta Bus length maximum length of the Modbus network in function of the Baud Rate It is the length of the cables that connect the two modules on which the bus terminal has been inserted by means of a DIPSwitch see Diagram 1 Derivation length maximum length of a derivation see Diagram 1 Diagram 1 Module 3 DL Module 1 Module 2 Module 5 DL Derivation length Bus length MODBUS Diagram 1 BUS length Deriva...

Страница 7: ...ION PARAMETERS 1 2 BAUD RATE 3 4 5 6 7 8 COMMUNICATION ADDRESS From software Communication parameters from EEPROM 9600 Baud Fixed address 01 19200 Baud Fixed address 02 38400 Baud Fixed address 03 57600 Baud Fixed address 04 X X X X X X Fixed address represented as binary Fixed address 63 Note if DIP switches from 1 to 8 are OFF the communication parameters are loaded from EEprom Sw3 MEASUREMENT A...

Страница 8: ...NPUT IN A IN B Max 1200 Vdc Max 850 Vac WARNING REFERRED TO GROUND Current output Voltage output 5 4 SW3 OFF 4 20 mA The output settings can be modified by the software 5 6 SW3 OFF 0 10 V The output settings can be modified by the software FACTORY SETTINGS Default conditions for module configuration parameters Full scale 1000 V Analogue output Current 4 20 mA Communication parameters 38400 8 N 1 A...