USER MANUAL – Z-SG
14
2) Switch Dip-Switch SW2-
1 as desired: “OFF”=digital input enabled, digital output disabled;
“ON”=digital input disabled, digital output enabled
3) Switch Dip-Switches SW2-2 and SW2-3 as desired: see Dip-Switches table
4) Switch Dip-Switches SW2-
4 to “OFF” and SW2-5 to “OFF”
5) Switch Dip-Switches SW2-
6 to “ON”, SW2-7 to “ON”, SW2-8 to “ON”
6) Power on the module
7) Write sensitivity value in reg. 40044, 40045 (FP)
8) Write load cell end scale in reg. 40046, 40047 (FP)
New sensitivity and load cell end scale are saved in Z-SG / Z-SG-L module.
10) Put the tare on the balance
11) Save the tare value in EEPROM memory (write 0xC2FA=49914 in reg.40068)
11.
MEASURE CALIBRATION WITHOUT MODBUS REGISTERS
11.1.
CALIBRATION WITH A STANDARD WEIGHT USING THE CALIBRATION
BUTTON
WARNING
Gross weight (tare + known weight) must not to exceed load cell end scale, to avoid
serious damage to the cell.
1) Power off the module before configuring it by Dip-Switches to avoid serious damage due to
electrostatic discharges.
2) Switch the Dip-Switches SW2-
4 to “ON” and SW2-5 to “ON”. In this way, setting by
calibration button is possible.
3) Switch the Dip-Switch SW2-
1 to “OFF”. In this way, calibration with known weight using
calibration button (or digital input) is possible.
4) Switch the Dip-Switches SW2-2 and SW2-3 as shown in Dip-Switches table, to select one of
the possible modalities of analog output.
5) Switch the Dip-Switches SW2-6, SW2-7, SW2-8 to choose the load cell sensitivity (see Dip-
Switch table)