SETTINGS BY INTERNAL JUMPERS AND TRIMMERS
Some jumpers and two trimmers are present on the board and allow to define some
custumized settings for the instrument.
To set the jumpers it is necessary to open the indicator from the rear side by applying
leverage using a screwdriver on the lateral joints and so draw the board.
The position of the internal jumpers within the board is illustrated on the below figure:
Input Type Settings
Input
0..20 mA
4..20 mA
0..5 V
1..5 V
0..10 V
2..10 V
Jumper P14
Closed
Opened
Opened
Opened
Closed
Opened
Opened
Closed
Opened
Closed
Opened
Opened
Opened
Closed
Opened
Closed
Opened
Opened
Closed
Opened
Opened
Opened
Closed
Closed
Jumper P15 Jumper P16 Jumper P18
Setting of the View Scale
To set the
o
display view, work on the jumper
:
scale f instantaneous
P5
- closed: full-scale of display 0..1000.
- opened: full-scale of display 1000..1999.
Connect a generator to the input of the instrument:
- provide the signal corresponding to the
(ex. 4 mA) and regulate the trimmer
zero-scale
TR2
up to obtain the desidered indication.
- provide the signal corresponding to the
(ex. 20 mA) and regulate the trimmer
full-scale
TR1
up to obtain the desidered indication.