 12

Battery Test

 

 
The LC103 has a built-in battery test feature that shows 
the remaining battery charge. A reading of 100% 
indicates that the battery is fully charged. As the battery 
charge is used up, the reading will drop. The low battery 
circuits will turn the unit off shortly after the battery test 
reading drops to 0%, and before the battery, level drops 
too low for reliable operation. The LC103 never fully 
discharges the battery, which helps extend the life of the 
BY289. 
 

To perform the battery test: 
1.

 

With a BY289 installed, press and hold the POWER 

button. 

2.

 

Read the graph and numerical percentage of 

remaining battery charge in the DISPLAY. 

3.

 

If the reading shows 0%, the unit may not operate, or 

operate for just a short time since the low battery 
circuit turns the LC103 off at this battery level. 

 
 
 

 

Fig. :  Press and hold the Power button to read the 
remaining battery charge.
 

 

Recharging the Battery 

 

The BY289 battery should never be allowed to remain 
discharged for more than a few hours, since this will 
shorten its lifetime. The battery must be recharged 
whenever the battery test reads 0%. However, you 
should recharge the battery more often than this to 
lengthen the battery's lifetime and keep the LC103 ready 
for portable use at all times. 

 

To recharge the battery, simply leave it installed inside 
the LC103 while the unit is connected to the AC Power 
Adapter and the Power Adapter is connected to a source 
of AC power. The charging time required to return the 
battery to 100% depends on how far it is discharged. The 
battery will trickle charge while the LC103 is in use and 
powered from the AC adapter, but it will recharge the 
quickest if the POWER is turned off. Normally, a battery 
will completely recharge in about 8 hours with the 
POWER off. 

To operate the LC103 using the optional BY289 
battery: 
1.

 

Install the BY289 battery into the LC103 battery 

compartment. 

NOTE:  If you are using the BY289 for the first time, 
be sure to charge the battery before using the LC103.  
Though factory tested, the BY289 may not be charged 
when you receive it. 

2.

 

Press the POWER button and release. The 

WARNING LED will momentarily blink to indicate 
it is operational and the displays will read “LC103 
ReZolver” and “Ready.” 

3.

 

The LC103 is immediately ready for use. If precise 

measurements are required, allow the unit to operate 
for 20 minutes to reach specified accuracy. 

 

Auto-Off 

 
To conserve battery charge, the LC103 contains an auto-
off circuit. This circuit keeps the batteries from running 
down if you should forget to turn the unit off, but keeps 
the ReZolver powered up during use. The auto-off 
circuit will shut the LC103 off after approximately 5  
 

 
 
 
minutes if none of the front panel buttons have been 
pushed. Pressing any COMPONENT TYPE button, 
COMPONENT PARAMETERS button, COMPONENT 
TEST button, or the POWER button resets the auto off 
circuits. The auto-off circuits are bypassed when the 
LC103 is operated from the AC Power Adapter. 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Содержание ReZolver LC103

Страница 1: ...PACITOR VALUE CAPACITOR ESR CAPACITOR LEAKAGE INDUCTOR RINGER DIELECTRIC ABSORPTION INDUCTOR VALUE V H mH H 5 6 7 8 9 0 CLR ALL OTHER CAPS SWITCHING TRANS FORMERS TANTALUM CAPS CERAMIC CAPS YOKES FLYB...

Страница 2: ...ilter capacitors after removing power before connecting to any part of the circuit requiring power to be removed 5 Be sure your equipment is in good order Broken or frayed test leads can be extremely...

Страница 3: ...1 LC103 ReZolver Capacitor Inductor Analyzer Operation and Application Manual 3200 Sencore Drive Sioux Falls SD 57107...

Страница 4: ...ng 25 Measuring Inductor Value 25 Automatic GOOD BAD Inductor Value Testing 26 Inductor Ringer Test 26 Error Messages 27 RS 232 OPERATION 28 Connecting to the Computer 28 Sending Data to the LC103 28...

Страница 5: ...Inductance Per Foot 45 Identifying Aging Cable 45 Testing High Voltage Diodes 46 High Potential Testing 46 Measuring High Voltage Resistors to 1 Gigaohm 47 APPENDIX Introduction 48 Capacitor Theory 4...

Страница 6: ...t of circuit capacitor and inductor analyzer It is designed to quickly identify defective capacitors and inductors by providing fast in circuit tests and complete analyzing out of circuit The test res...

Страница 7: ...nged 0 01 uA 0 01 uA to 19 99 uA 0 l uA 20 0 uA to 199 9 uA 1 0 uA 200 uA to 1999 uA 0 01 mA 2 00 mA to 19 99 mA EQUIVALENT SERIES RESISTANCE ESR U S Patent 4 795 966 ACCURACY 5 l digit CAPACITOR RANG...

Страница 8: ...nute warm up 50o to 86o F 10o to 30o C HUMIDITY 0 90 no condensation MAXIMUM ALTITUDE 4 000 meters POWER 105 130V AC 60Hz 30 watts with supplied PA251 power adapter Battery operation with optional BY2...

Страница 9: ...circuit 11 OUT OF CIRCUIT Test Buttons a Capacitor GOOD BAD Press to measure capacitor value and ESR in circuit b Inductor GOOD BAD Press to measure inductance value in circuit c Dielectric Absorption...

Страница 10: ...SUPPLIED ACCESSORIES 1 Out of Circuit Test Leads 39G219 Special low capacity cable with E Z Hook clips Connect to TEST LEAD jack 2 Adjustable In Circuit Test Probe AP291 patent pending Exclusive in ci...

Страница 11: ...es 4 Field Calibrator FC221 Use to periodically check the calibration of the LC103 5 Rechargeable Battery BY289 Lead acid battery provides portable operation for the LC103 6 SCR TRIAC Test Accessory S...

Страница 12: ...on low value of components ATTENTION Utilisez seulement le PA251 au PA252 de Sencore comme adaptateur secteur L utilisation d un adaptateur secteur pour le LC103 diff rent du PA251 ou du PA252 peut e...

Страница 13: ...ns when using lead acid batteries 1 Do not dispose of old lead acid batteries in fire This may cause them to burst spraying acid through the air 2 Do not short the and terminals together This will bur...

Страница 14: ...ter and the Power Adapter is connected to a source of AC power The charging time required to return the battery to 100 depends on how far it is discharged The battery will trickle charge while the LC1...

Страница 15: ...counterclockwise until the fuse releases 3 Replace the fuse with the same type and rating 1 amp 2AG Slo Blo 4 Use the screwdriver to gently push inward while turning the fuse holder clockwise back int...

Страница 16: ...lead mounting clip holds the test leads out of the way yet ready for use at anytime NOTE Mounting the TEST LEAD MOUNTING CLIP to the sides of the ReZolver will interfere with the handle movement Lead...

Страница 17: ...ample 123456 b The LC103 accepts numbers up to 4 places after the decimal Example 1 2345 c Push the CLR button once to clear the value entry and start over 2 Enter the desired capacitor value multipli...

Страница 18: ...OMPONENT TEST RESULTS CAPACITOR INDUCTOR ANALYZER POWER HOLD FOR BATTERY TEST ALUMINUM LYTICS DOUBLE LAYER LYTICS HIGHER DOUBLE LAYER 1 2 3 4 pF F F CAPACITOR GOOD BAD INDUCTOR GOOD BAD CAPACITOR VALU...

Страница 19: ...ote The TEST ACTIVATION button on the probe can be used if a capacitor COMPONENT TYPE is selected Automatic EIA GOOD BAD Testing To perform the automatic EIA GOOD BAD test simply enter the capacitors...

Страница 20: ...be entered to perform any out of circuit capacitor parameter test If any blue Inductor Component Type button is selected Component Type Selection Error will appear in the display when you attempt to m...

Страница 21: ...e percentage of dielectric absorption on the COMPONENT TEST RESULTS display Standard Aluminum Electrolytic Capacitors Capacity in uF 1 5V 3 0V 6 0V 10V 15V 20V 25V 35V 50V 100V 200V 300V 400V 500V 600...

Страница 22: ...0 02 0 02 0 02 0 02 0 02 0 02 0 02 0 02 4700 0 03 0 03 0 03 0 03 0 03 0 03 0 03 0 02 0 02 0 02 0 02 0 02 0 02 0 02 0 02 0 02 6800 0 02 0 02 0 02 0 02 0 02 0 02 0 02 0 01 0 01 0 01 0 01 0 01 0 01 0 01...

Страница 23: ...ACITOR and the STOP TESTING alarm may activate until the voltage is gone from the test leads All data input and test buttons will be locked out until the display returns to Ready Leakage In Paper Mica...

Страница 24: ...3 3000 3550 4243 6000 8485 15000 900 1273 1800 2324 2846 3286 3674 4347 5196 7348 22000 1090 1541 2180 2814 3447 3980 4450 5265 6293 8899 33000 1335 1888 2671 3447 4221 4874 5450 6448 7707 47000 1593...

Страница 25: ...0 25 28 32 38 45 64 90 110 127 142 156 201 470 24 24 24 24 29 34 38 45 54 76 107 131 152 170 186 240 680 29 29 29 29 35 41 46 54 65 91 129 158 183 204 224 289 1000 35 35 35 35 43 49 55 65 78 111 157 1...

Страница 26: ...e capacitor to be tested The LC103 determines the GOOD BAD limits If you desire to grade capacitors according to value you must also enter the desired and value tolerances The value tolerances however...

Страница 27: ...measures the true inductance of coils using a fast reliable patented test Coils from 0 l uH to 19 99 H are automatically measured for value by connecting the test leads and pressing the test button A...

Страница 28: ...n yokes horizontal flyback transformers and switching power supply transformers are reliably checked with the Ringer test The LC103 automatically matches the coil impedance to the necessary testing pa...

Страница 29: ...ameter entered via the keypad or RS 232 is beyond the limits of the automatic GOOD BAD test The component may still be able to be tested but not for a GOOD BAD indication Possible causes 1 Performing...

Страница 30: ...The LC103 is configured as a DCE Data Communications Equipment device which means that it should be connected to your computer with a cable having DB9 connectors on both ends and straight through wir...

Страница 31: ...ree are only used for LC103 automatic GOOD BAD comparisons The capacitor voltage code also sets the LC103 power supply to the selected voltage for the leakage test When sending a component value to th...

Страница 32: ...ds when the computer sends any other listener code or when the front panel button is pressed WARNING The warning LED on the front of the LC103 will falsh as a reminder that a shock hazard of up to 100...

Страница 33: ...e NVC ou une autre commande afin de vider la function de test Le LC103 peut tre endommag si une capacit charg e ou une tension statique lui est connect e De plus l utilisanteur peut recevoir une d cha...

Страница 34: ...g up waiting to send its last linefeed character If this happens you may need to put an extra GET or INPUT statement into your program to let the LC103 send its last character into an unused variable...

Страница 35: ...ther case the computer can base its decision on whether the capacitor is good or bad by using the GOOD BAD Indicator in the returned data For the automatic GOOD BAD test to function the capacitor s va...

Страница 36: ...type to be used When troubleshooting a circuit it is not important to know why a certain type of capacitor was selected It is best to simply replace a bad capacitor with a good capacitor of the same t...

Страница 37: ...same value and voltage rating While they may have many shapes tantalum capacitors always have polarized leads Lead polarization is often the only way to distinguish a tantalum lytic from another type...

Страница 38: ...tor types are the axial lead and chip types Some axial lead ceramic capacitors may look the same as resistors and inductors which also use the same case type You can easily determine if the component...

Страница 39: ...tify They too are used exclusively in video applications and produce high voltage for the CRT A flyback has several terminals which are often soldered to a PC board chassis One or two heavily insulate...

Страница 40: ...fy by physical appearance or any other means CAPACITOR TESTING Capacitance Measurement Accuracy The LC103 measures the capacitor charge time as the capacitor is charged with a constant current This gi...

Страница 41: ...other capacitors in parallel It may be necessary to remove the suspect capacitor to test it accurately 2 Sometimes the LC103 will return a value and ESR reading accompanied by SUGGEST REMOVAL In this...

Страница 42: ...APACITOR LEAKAGE button and read the amount of leakage on the display It must be within the maximum allowable leakage limits for its value and voltage rating 4 Connect one end of a short jumper to the...

Страница 43: ...of the ring energy A shielded coil is good if it rings ten or more However if it rings less than ten remove the metal shield if possible and test the coil again If it now rings 10 or more the coil is...

Страница 44: ...them are shorted or leaky 5 If all the loads are disconnected and the ringing test still indicates BAD completely remove it from the circuit If the flyback primary still rings less than 10 the flybac...

Страница 45: ...the INDUCTOR RINGER test button and read the test result in the LC103 display 6 If the ringer test result is 10 or more the horizontal yoke is good If it rings less than 10 the horizontal yoke is defe...

Страница 46: ...lso use this test to determine the length or to pinpoint a break in multi conductor cable that has 3 or more conductors Due to variations in conductor spacing and noise pickup however the accuracy wil...

Страница 47: ...they need to be replaced The LC103 can be used for preventative maintenance checks of coaxial cable to determine if deterioration is beginning to occur As a cable begins to fail the dielectric separa...

Страница 48: ...ny higher voltage as this will cause excessive current flow through the diode and damage it 5 If you apply 999 9 volts to the diode and it still shows no conduction it is open and you do not need to c...

Страница 49: ...will read the resistance of these resistors without any calculations The range of resistance which the LC103 will measure depends on the applied voltage Table 13 shows the amount of applied voltage n...

Страница 50: ...the dipoles are stressed and forced to line up in rows creating stored energy in the dielectric The dielectric has undergone a physical change similar to that of soft iron when exposed to current thro...

Страница 51: ...failure condition for aluminum lytic capacitors Tantalum Electrolytics The tantalum electrolytic capacitor is also quite popular While the leakage in the aluminum lytic is very high due to the nature...

Страница 52: ...perature change versus capacity change of non temperature compensated ceramic disc capacitors 5 0 4 0 3 0 2 0 1 0 0 1 0 2 0 3 0 4 0 55 45 35 25 15 5 5 15 25 35 45 55 65 75 85 95 105 115 125 Temperatur...

Страница 53: ...are another example of capacitors that can change value in circuit or on the shelf As these capacitors dry out they eventually lose their capacitance due to the failure of the aluminum oxide film maki...

Страница 54: ...d to a capacitor a certain amount of current will flow through the capacitor This current is called the leakage current and is the result of imperfections in the dielectric Whenever this leakage curre...

Страница 55: ...53...

Страница 56: ...54...

Страница 57: ...55...

Страница 58: ...56...

Страница 59: ...charge time is expressed in a percentage of the original charge Also called Capacitor Memory or Battery Action Dielectric constant the ratio of capacitance between a capacitor having a dry air dielect...

Страница 60: ...mic disc capacitor Derived from negative positive zero Does not change capacity with temperature changes Padder a high capacity variable capacitor placed in series with a fixed capacitor to vary the t...

Страница 61: ...__________________________________________________________________ __________________________________________________________________________ __________________________________________________________...

Страница 62: ...__________________________________________________________________ __________________________________________________________________________ __________________________________________________________...

Страница 63: ...ections 1 Use a corrugated cardboard shipping container that has a test strength of 32 lbs inch ECT 2 Call 1 800 SENCORE and ask for a return unit authorization RUA number Enclose the following inform...

Страница 64: ...62 3200 Sencore Drive Sioux Falls SD 57107 Call 1 800 SENCORE 736 2673 Innovatively designed www sencore com with your time in mind Form6808A Printed in the U S A...

Отзывы: