13
Function
Parameter Settings/Options
Initial Zero Range
C21
0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
Zero Tracking
C22
0= turn off zero tracking
0.5 = ±0.5d d = division
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d
Note: the zero tracking range can not be bigger than manual zero
range
Zero Tracking Time
C23
0 = turn off zero tracking time
1 = 1 second
2 = 2 seconds
3 = 3 seconds
Overload Range
C24
00 = turn off overload range
01-99d = overload range setting
d = division
Negative Display
C25
0 = -9d
10 = -10% max. capacity
20 = -20% max. capacity
50 = -50% max. capacity
100 = -100% max. capacity
Standstill Time
C26
0 = quick
1 = medium
2 = slow
Standstill Range
C27
1 = ±1d
d= division
2 = ±2d
5 = ±5d
10 = ±10d
Digital Filter
(for filtering moving
weight, such as
animals)
C28
0 = turn off dynamic filter
1 = Low dynamic filter
3 = Medium dynamic filter
5 = High dynamic filter
Noise Filter
C29
0 = turn off noise filter
1 = Low
2 = Medium
3 = High
Print Time and Date
C30
0 = yy.mm.dd
1 = mm.dd.yy
2 = dd.mm.yy
3 = yy.mm.dd
Gravity of Calibration
Location
C36
9.7000 - 9.9999
Gravity of Destination
C37
9.7000 - 9.9999
Version No.
C38
Input Signal
C39
0 = Input Signal Direction Normal
1 = Input Signal Direction Reversed