8

Hold

In the parameter settings you can choose one of these 5 hold options

1. Peak Hold: Grabs the highest weight 

(for materials testing, ie. tension and pulling

force)

● Press the HOLD key then add weight to the scale

● The indicator will show the highest weight it recorded and hold it on the screen until

a higher weight is placed on the scale

2. Manual Hold: Grabs the current weight and holds it so it will not change/fluctuate

● While weighing, press HOLD and the indicator will hold the current weight on the

screen until HOLD is pressed again

3. Auto Hold: If the weight on the scale is above 20d and is stable, the indicator will

hold that weight on the screen for 3 seconds then go back to general weighing

● Pressing the hold key is unnecessary, holding is done automatically when the scale

is stable

4. Average Hold: Used for animal weighing, the indicator will display the average

weight sampled from 3 or 5 seconds (Set in the C12 Parameter)

● Add animal to scale and press HOLD
● Indicator screen will show “

LOC

” for 3 or 5 seconds, then display the average

weight from those 3 seconds

● Press HOLD again to exit holding mode

5. Auto Average Hold: Used for animal weighing, the indicator will display the aver-

age weight sampled from 3 or 5 seconds without the need to press the HOLD key.

If the weight on the scale is above 20d and is stable, the indicator will start grabbing

the average weight sampled from 3 seconds.

● Begin loading animal to scale, after 5 seconds the Indicator screen will show “

LOC

for 3 seconds

● It will then display the average weight from those 3 seconds for 3 seconds and then

repeat the process

Print

● If the indicator is connected to a printer and the communication mode is set to print

mode (C18 =2) and the weight on the scale is stable press and hold the SET key to
print the current weight

Print out example:

 

 

  Net 

weight

 

 

  Tare 

weight

 

 

  Gross 

weight

 

    Print 

out 

number

N.W.:

25.6lb

T.W.:

10.3lb

G.W.:

35.9lb

NO.

01

Содержание SL-7516 Series

Страница 1: ...www selletonscales com SL 7516 SERIES INDICATOR USER S MANUAL SL 7516 E SL 7516 C Series...

Страница 2: ...Set Up 1 Features 2 Specifications 3 Power Supply 4 Displays 5 Display and Key Descriptions 6 Operating Instructions 7 8 Calibration 9 10 Indicator Parameter Settings 11 13 Definitions 14 Connectors 1...

Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Страница 4: ...s Accuracy class 3000 e Stimulating voltage 3 3 VDC A D converting speed 10 SPS Load signal range 0 12 8mV Load capacity can connect 4 pcs 350 load cell at most Interval 1 2 5 10 20 50 Display 6 digit...

Страница 5: ...3 6in 8 7in 6in 3 2in power supply load cellport batterybox rs232 port SPECIFICATIONS INDICATOR MEASUREMENTS...

Страница 6: ...use for 10 12 hours to prevent low voltage resulted from self leakage of batter Once charged the battery should last for 45 hours To keep the battery in best condition fully discharge the battery eve...

Страница 7: ...5 SL 7516 E LED SL 7516 C LCD...

Страница 8: ...th the On Off button to enter and exit calibration The scale is at zero The scale is stable Gross Shows you are in Gross weight mode includes tare default mode Net Shows you are in Net weight mode wit...

Страница 9: ...to tare out the weight Unit Selection To switch between measuring units kg lb oz press the UNITS key Tare Function The Tare function is used when you only wish to see the current change in weight not...

Страница 10: ...display the average weight sampled from 3 or 5 seconds Set in the C12 Parameter Add animal to scale and press HOLD Indicator screen will show LOC for 3 or 5 seconds then display the average weight fr...

Страница 11: ...this scale by using HOLD and UNITS to move the cursor left and right and TARE and ZERO to move the values down and up The C4 channel is used to enter in the max capacity of the scale Make sure this do...

Страница 12: ...calibration weight on the scale 30 If the scale does not show the value of the calibration weight check that the feet on the platform are not screwed in too tightly and verify that the platform is le...

Страница 13: ...ter key as labeled under each indicator button 3 Press the SET key to enter edit the parameter setting 4 Press the ON OFF key to save and exit settings at any time Table 1 Calibration Parameter Settin...

Страница 14: ...utomatically holds data when stable 4 Average hold for animal weighing averages the weight from a sample of 3 or 5 seconds Set in parameter C12 5 Auto Average hold Average hold without the need to pre...

Страница 15: ...range setting d division Negative Display C25 0 9d 10 10 max capacity 20 20 max capacity 50 50 max capacity 100 100 max capacity Standstill Time C26 0 quick 1 medium 2 slow Standstill Range C27 1 1d...

Страница 16: ...the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the set calibrated range Adds a tolerance...

Страница 17: ...wire load cell connections are both okay Please contact us directly if you have other special needs for your application Quick Disconnect as shown below QUICK DISCONNECT CONNECTION DIAGRAM DB9 SERIAL...

Страница 18: ...t Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overload S2 weight mode GS gross mode NT net mode S3 weight of...

Страница 19: ...e the max capacity Use correct weight within the defined range ERR2 During calibration the weight is below the minimum required weight The calibration weight minimum is 10 of the max capacity set in C...

Отзывы: