SellEton Scales SL-7516 Series Скачать руководство пользователя страница 14

12

Function

Parameter Settings/Options

Restore Default

Settings

C07

0 = do not restore

1 = restore to default settings

Warning Tone

C08

0 = turn off warning tone

1 = turn on warning tone

Power Off  

Automatically

C09

0 = turn off auto power off

10 = power off automatically if no change within 10 minutes

30 = power off automatically if no change within 30 minutes

60 = power off automatically if no change within 60 minutes

Power Saving Mode

C10

LED Version OP901A:

0 = turn off power saving setting

3 = turn off display if no change within 3 minutes

5 = turn off display if no change within 5 minutes

LCD Version OP901B:

0 = turn off the backlight 

1 = backlight only when the weight changes or keyboard is pressed

2 = constant backlight

Hold Function

C11

0 = turn off hold function

1 = Peak hold - Grabs the highest weight

2 = Manual hold - Grabs the current weight

3 = Auto hold - Automatically holds data when stable

4 = Average hold - for animal weighing, averages the weight from a    

      sample of 3 or 5 seconds (Set in parameter C12)

5 = Auto Average hold - Average hold without the need to press the 

      hold key

Hold Time

C12

If you chose C11=4 this setting allows you to set how many 

seconds it samples to obtain an average weight

3 = 3 seconds

5 = 5 seconds

Upper Limit Alarm

C13

Set upper limit within the max. capacity

Lower Limit Alarm

C14

Set lower limit within the max. capacity

Inner Code Display

C15

Check the inner code (raw data)

Communication 

Setting

C18

Set the serial interface data output method:

0 = Turn off serial interface data output

1 = Continuous sending mode, connect 

remote display

2 = Print mode, connect printer

3 = n/a

4 = PC continuous sending mode, connect computer

Baud Rate

C19

0=1200 (for remote display)

1=2400

2=4800

3=9600

Manual

Zero Range

C20

0 = turn off manually zero setting

1 = ±1% max capacity

2 = ±2% max capacity

4 = ±4% max capacity

10 = ±10% max capacity

20 = ±20% max capacity

100 = ±100% max capacity

Table 2. Indicator Parameter Settings

Содержание SL-7516 Series

Страница 1: ...www selletonscales com SL 7516 SERIES INDICATOR USER S MANUAL SL 7516 E SL 7516 C Series...

Страница 2: ...Set Up 1 Features 2 Specifications 3 Power Supply 4 Displays 5 Display and Key Descriptions 6 Operating Instructions 7 8 Calibration 9 10 Indicator Parameter Settings 11 13 Definitions 14 Connectors 1...

Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Страница 4: ...s Accuracy class 3000 e Stimulating voltage 3 3 VDC A D converting speed 10 SPS Load signal range 0 12 8mV Load capacity can connect 4 pcs 350 load cell at most Interval 1 2 5 10 20 50 Display 6 digit...

Страница 5: ...3 6in 8 7in 6in 3 2in power supply load cellport batterybox rs232 port SPECIFICATIONS INDICATOR MEASUREMENTS...

Страница 6: ...use for 10 12 hours to prevent low voltage resulted from self leakage of batter Once charged the battery should last for 45 hours To keep the battery in best condition fully discharge the battery eve...

Страница 7: ...5 SL 7516 E LED SL 7516 C LCD...

Страница 8: ...th the On Off button to enter and exit calibration The scale is at zero The scale is stable Gross Shows you are in Gross weight mode includes tare default mode Net Shows you are in Net weight mode wit...

Страница 9: ...to tare out the weight Unit Selection To switch between measuring units kg lb oz press the UNITS key Tare Function The Tare function is used when you only wish to see the current change in weight not...

Страница 10: ...display the average weight sampled from 3 or 5 seconds Set in the C12 Parameter Add animal to scale and press HOLD Indicator screen will show LOC for 3 or 5 seconds then display the average weight fr...

Страница 11: ...this scale by using HOLD and UNITS to move the cursor left and right and TARE and ZERO to move the values down and up The C4 channel is used to enter in the max capacity of the scale Make sure this do...

Страница 12: ...calibration weight on the scale 30 If the scale does not show the value of the calibration weight check that the feet on the platform are not screwed in too tightly and verify that the platform is le...

Страница 13: ...ter key as labeled under each indicator button 3 Press the SET key to enter edit the parameter setting 4 Press the ON OFF key to save and exit settings at any time Table 1 Calibration Parameter Settin...

Страница 14: ...utomatically holds data when stable 4 Average hold for animal weighing averages the weight from a sample of 3 or 5 seconds Set in parameter C12 5 Auto Average hold Average hold without the need to pre...

Страница 15: ...range setting d division Negative Display C25 0 9d 10 10 max capacity 20 20 max capacity 50 50 max capacity 100 100 max capacity Standstill Time C26 0 quick 1 medium 2 slow Standstill Range C27 1 1d...

Страница 16: ...the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the set calibrated range Adds a tolerance...

Страница 17: ...wire load cell connections are both okay Please contact us directly if you have other special needs for your application Quick Disconnect as shown below QUICK DISCONNECT CONNECTION DIAGRAM DB9 SERIAL...

Страница 18: ...t Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overload S2 weight mode GS gross mode NT net mode S3 weight of...

Страница 19: ...e the max capacity Use correct weight within the defined range ERR2 During calibration the weight is below the minimum required weight The calibration weight minimum is 10 of the max capacity set in C...

Отзывы: