![SellEton Scales SL-7516 Series Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/selleton-scales/sl-7516-series/sl-7516-series_user-manual_1237068014.webp)
12
Function
Parameter Settings/Options
Restore Default
Settings
C07
0 = do not restore
1 = restore to default settings
Warning Tone
C08
0 = turn off warning tone
1 = turn on warning tone
Power Off
Automatically
C09
0 = turn off auto power off
10 = power off automatically if no change within 10 minutes
30 = power off automatically if no change within 30 minutes
60 = power off automatically if no change within 60 minutes
Power Saving Mode
C10
LED Version OP901A:
0 = turn off power saving setting
3 = turn off display if no change within 3 minutes
5 = turn off display if no change within 5 minutes
LCD Version OP901B:
0 = turn off the backlight
1 = backlight only when the weight changes or keyboard is pressed
2 = constant backlight
Hold Function
C11
0 = turn off hold function
1 = Peak hold - Grabs the highest weight
2 = Manual hold - Grabs the current weight
3 = Auto hold - Automatically holds data when stable
4 = Average hold - for animal weighing, averages the weight from a
sample of 3 or 5 seconds (Set in parameter C12)
5 = Auto Average hold - Average hold without the need to press the
hold key
Hold Time
C12
If you chose C11=4 this setting allows you to set how many
seconds it samples to obtain an average weight
3 = 3 seconds
5 = 5 seconds
Upper Limit Alarm
C13
Set upper limit within the max. capacity
Lower Limit Alarm
C14
Set lower limit within the max. capacity
Inner Code Display
C15
Check the inner code (raw data)
Communication
Setting
C18
Set the serial interface data output method:
0 = Turn off serial interface data output
1 = Continuous sending mode, connect
remote display
2 = Print mode, connect printer
3 = n/a
4 = PC continuous sending mode, connect computer
Baud Rate
C19
0=1200 (for remote display)
1=2400
2=4800
3=9600
Manual
Zero Range
C20
0 = turn off manually zero setting
1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
Table 2. Indicator Parameter Settings