SellEton Scales SL-7515 Скачать руководство пользователя страница 9

7

CALIBRATION cont.

If FS = 6

Please follow the procedure below: 

The display will show [

dP 1

], to indicate how many decimal places you want;

press UNIT to change, and press HOLD to confirm (ex. DP  2 = ####.##)

The display will show [

d  2

], to indicate the division; press UNIT to change, and

press HOLD to confirm

The display will show [

CAP 

], to indicate the max. capacity; press HOLD to

display the current max. capacity; press ZERO to left shift and UNIT to change the
capacity; Then press HOLD to confirm

The display will show [

LoAd 

], to indicate the calibration loading weight; press

HOLD to display the current calibration loading weight; press ZERO to left shift and

UNIT to change the calibration loading weight; Then press HOLD to confirm

The display will show [

CALon

] please follow the steps for FS values  of 0-6 to

complete calibration

For FS values of 0 - 6

Please follow the procedure below: 

The display will show [

CALon

], press HOLD to enter into weight calibraion, or

press ON/OFF to exit calibration

The display will show [

noLoAd

] for zero calibration; make sure there is no load

on scale, then press HOLD to confirm and zero calibrate

The display will show [

AdLoAd

] for loading calibration; place calibration weight

as defined by the tables above on the scale

● Once the weighing platform is stable, press HOLD to confirm and calibrate

The display will show [

CALEnd

] when the calibration is complete, press HOLD to

save the weight calibration and press ON/OFF to exit and go back to weighing mode

Содержание SL-7515

Страница 1: ...www selletonscales com SL 7515 INDICATOR USER S MANUAL...

Страница 2: ...1 Features 2 Technical Parameters 2 Power Supply 2 Display and Key Descriptions 3 Operating Instructions 4 Calibration 5 6 Indicator Parameter Settings 7 Helpful Definitions 8 Troubleshooting 8 Conne...

Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Страница 4: ...l zero range 10 max capacity Manual zero range 2 max capacity Resolution Display Up to 1 15 000 Display Resolution Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70...

Страница 5: ...d Confirm ZERO Zero s the scale Tare s the scale Move Left UNITS Shifts between weighing units kg lb Move Down DISPLAY Meaning The scale is at zero The scale is stable Gross Shows you are in Gross wei...

Страница 6: ...ot at gross zero due to material build up If the weight on the scale is within the zero range 2 of the max capacity and stable and ZERO is pressed the indicator will zero the scale and stay in gross w...

Страница 7: ...nimal weighing the indicator will display the average weight sampled from 1 9 seconds Set in the HAt parameter setting Add animal to scale and press HOLD Indicator screen will show LOC for 1 9 seconds...

Страница 8: ...low for other variances Press UNIT to change the capacity range to your scales capacity Calibration chart for kilogram usage Display Max Capacity Division Calibration Loading Weight FS 0 15 00kg 0 005...

Страница 9: ...ift and UNIT to change the calibration loading weight Then press HOLD to confirm The display will show CALon please follow the steps for FS values of 0 6 to complete calibration For FS values of 0 6 P...

Страница 10: ...ample of 3 seconds 5 Auto Average hold Average hold without pressing HOLD Hold Time Ht 1 0 turn off hold time 1 15 second hold time 2 30 second hold time 3 45 second hold time 4 60 second hold time 5...

Страница 11: ...9 Function Display Parameter Settings Baud Rate bAud 1 1 9600bit s 2 4800bit s 3 2400bit s 4 1200bit s 5 600bit s...

Страница 12: ...erred in a communication channel example In the serial port context 9600 baud means that the serial port is capable of transferring a maximum of 9600 bits per second TROUBLESHOOTING Error Reason Solut...

Страница 13: ...uick Disconnect DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes depending on the indicator model Figure 4 shows the pin assignment on the DB9 9 pin...

Страница 14: ...2 port and the indicator will respond with the current data Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overl...

Страница 15: ...IFICATIONS CONTACT US 2 5 2 ON OFF HOLD UNIT ZERO 8 4 5 Please e mail info selletonscales com for any sales related questions or call 844 735 5386 Don t forget to visit our website at www selletonscal...

Отзывы: