SellEton Scales SL-7515 Скачать руководство пользователя страница 10

8

PARAMETER SETTINGS

Press HOLD and UNIT together to enter the
parameter settings

Use the following diagram to navigate through
the parameter settings:

Function

Display

Parameter Settings

Automatic

 

Power Off

OFF 10

0 = turn off auto power off

10 = power off automatically if no change within 10 minutes

30 = power off automatically if no change within 30 minutes

60 = power off automatically if no change within 60 minutes

90 = power off automatically if no change within 90 minutes

Power Saving Mode

BL   1

0 = turn off the backlight 

1 = backlight only when the weight changes or keyboard is pressed

2 = constant backlight

Unit Conversion

UnS  1

0 = turn off unit switch function

1 = kg/lb

2 = lb/oz

3 = kg/lb/oz

Hold Function

Hold 1

0 = turn off hold function

1 = Peak hold - Grabs the highest weight

2 = Manual hold - Grabs the current weight

3 =

 

Auto hold - Automatically holds data when stable 

4 = Average hold - Averages the weight from a sample of 3 seconds

5 = Auto Average hold - Average hold without pressing HOLD

Hold Time

Ht   1

0 = turn off hold time

1 = 15 second hold time

2 = 30 second hold time

3 =

 45 second hold time

4 = 60 second hold time

5 = 90 second hold time

Average Hold 

Sample Time

HAt  1

1-9 Seconds = How many seconds sampled for the average weight

Loading Time

HAdt 1

1-9 Seconds = How long the scale waits (So you can load the ani

-

mal on the scale) until  it begins to take the average weight

Zero Range

(% scale is allowed 

to zero, anything 

after will be tared)

2Ero 1

0 = turn off manually zero setting

1 = Initial zero range ±10% max capacity,

      Manual zero range ±2%

2 = Initial zero range ±100% max capacity,

      Manual zero range ±100%

Negative Range

(% it can go nega-

tive before an error)

nEgr 1

0 = -20d negative display

1 = 10% MAX negative display

2 = 100% MAX negative display

Communication 

Setting

trn  1

Set the serial interface data output method:

0 = Turn off serial interface data output

1 = Continuous sending mode, connect computer

2 = Print mode, connect printer

3 = Command request mode, connect computer

KEY

Function

ON/OFF

Exit

HOLD

Confirm

ZERO

Move Left

UNITS

Move Down

Содержание SL-7515

Страница 1: ...www selletonscales com SL 7515 INDICATOR USER S MANUAL...

Страница 2: ...1 Features 2 Technical Parameters 2 Power Supply 2 Display and Key Descriptions 3 Operating Instructions 4 Calibration 5 6 Indicator Parameter Settings 7 Helpful Definitions 8 Troubleshooting 8 Conne...

Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Страница 4: ...l zero range 10 max capacity Manual zero range 2 max capacity Resolution Display Up to 1 15 000 Display Resolution Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70...

Страница 5: ...d Confirm ZERO Zero s the scale Tare s the scale Move Left UNITS Shifts between weighing units kg lb Move Down DISPLAY Meaning The scale is at zero The scale is stable Gross Shows you are in Gross wei...

Страница 6: ...ot at gross zero due to material build up If the weight on the scale is within the zero range 2 of the max capacity and stable and ZERO is pressed the indicator will zero the scale and stay in gross w...

Страница 7: ...nimal weighing the indicator will display the average weight sampled from 1 9 seconds Set in the HAt parameter setting Add animal to scale and press HOLD Indicator screen will show LOC for 1 9 seconds...

Страница 8: ...low for other variances Press UNIT to change the capacity range to your scales capacity Calibration chart for kilogram usage Display Max Capacity Division Calibration Loading Weight FS 0 15 00kg 0 005...

Страница 9: ...ift and UNIT to change the calibration loading weight Then press HOLD to confirm The display will show CALon please follow the steps for FS values of 0 6 to complete calibration For FS values of 0 6 P...

Страница 10: ...ample of 3 seconds 5 Auto Average hold Average hold without pressing HOLD Hold Time Ht 1 0 turn off hold time 1 15 second hold time 2 30 second hold time 3 45 second hold time 4 60 second hold time 5...

Страница 11: ...9 Function Display Parameter Settings Baud Rate bAud 1 1 9600bit s 2 4800bit s 3 2400bit s 4 1200bit s 5 600bit s...

Страница 12: ...erred in a communication channel example In the serial port context 9600 baud means that the serial port is capable of transferring a maximum of 9600 bits per second TROUBLESHOOTING Error Reason Solut...

Страница 13: ...uick Disconnect DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes depending on the indicator model Figure 4 shows the pin assignment on the DB9 9 pin...

Страница 14: ...2 port and the indicator will respond with the current data Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overl...

Страница 15: ...IFICATIONS CONTACT US 2 5 2 ON OFF HOLD UNIT ZERO 8 4 5 Please e mail info selletonscales com for any sales related questions or call 844 735 5386 Don t forget to visit our website at www selletonscal...

Отзывы: