HELPFUL DEFINITIONS

Division:

 The amount of increments a scale offers. How accurate the scale can be

Capacity: the maximum amount the scale can contain

Zero Range: % of the max capacity that the scale is allowed to zero, anything above 

this percentage number will be tared.

Initial Zero Range: The percentage of weight allowed on the scale when indicator is powered 

on that will automatically zero (anything above this percent will be tared). 

example: If initial zero range is set to 10% of the max. capacity and your max. capacity is 

100lbs, you can place up to 10lbs of weight on the scale and when the indicator is powered on, it 

will automatically zero out the weight. 

Manual Zero Range: The percentage of weight allowed on the scale where the indicator will let 

you manually zero by pressing the ZERO key (anything above this percent will be tared).

Negative Range: How far you can go in the negative direction before displaying an 
error code

Baud Rate:

 The rate at which information is transferred in a communication channel. 

example: In the serial port context, “9600 baud” means that the serial port is capable of 
transferring a maximum of 9600 bits per second.

TROUBLESHOOTING

Error

Reason

Solution

UUUUUU

1. Overload

2. Wrong connection with load cell

3. Load cell has quality problem

1. Reduce the weight

2. Check load cell connection

3. Inspect load cell; Check the input/output

nnnnnnn

1. Calibration is no good

2. Wrong connection with load cell

3. Load cell has quality problem

1. Make sure scale is level

2. Check load cell connection

3. Check load cell input and output resistance

ERR1

During calibration, weight is not used or 

the weight is above the max. capacity

Use correct weight within the defined range

ERR2

During calibration, the weight is below 

the minimum required weight

The calibration weight minimum is 10% of 

the max. capacity set in C04.

 

Recommended to use 60%-80% of max.

 

capacity if possible

ERR3

During calibration, the input signal is 

negative

1. Check all wire connections

2. Check load cell

3. Recalibrate

4. PCB replacement needed if steps 1-3 fail

ERR4

During calibration signal is unstable

After the platform is stable, start calibration

ERR5

EEPROM Error

Change PCB

Содержание SL-7515

Страница 1: ...www selletonscales com SL 7515 INDICATOR USER S MANUAL...

Страница 2: ...1 Features 2 Technical Parameters 2 Power Supply 2 Display and Key Descriptions 3 Operating Instructions 4 Calibration 5 6 Indicator Parameter Settings 7 Helpful Definitions 8 Troubleshooting 8 Conne...

Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...

Страница 4: ...l zero range 10 max capacity Manual zero range 2 max capacity Resolution Display Up to 1 15 000 Display Resolution Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70...

Страница 5: ...d Confirm ZERO Zero s the scale Tare s the scale Move Left UNITS Shifts between weighing units kg lb Move Down DISPLAY Meaning The scale is at zero The scale is stable Gross Shows you are in Gross wei...

Страница 6: ...ot at gross zero due to material build up If the weight on the scale is within the zero range 2 of the max capacity and stable and ZERO is pressed the indicator will zero the scale and stay in gross w...

Страница 7: ...nimal weighing the indicator will display the average weight sampled from 1 9 seconds Set in the HAt parameter setting Add animal to scale and press HOLD Indicator screen will show LOC for 1 9 seconds...

Страница 8: ...low for other variances Press UNIT to change the capacity range to your scales capacity Calibration chart for kilogram usage Display Max Capacity Division Calibration Loading Weight FS 0 15 00kg 0 005...

Страница 9: ...ift and UNIT to change the calibration loading weight Then press HOLD to confirm The display will show CALon please follow the steps for FS values of 0 6 to complete calibration For FS values of 0 6 P...

Страница 10: ...ample of 3 seconds 5 Auto Average hold Average hold without pressing HOLD Hold Time Ht 1 0 turn off hold time 1 15 second hold time 2 30 second hold time 3 45 second hold time 4 60 second hold time 5...

Страница 11: ...9 Function Display Parameter Settings Baud Rate bAud 1 1 9600bit s 2 4800bit s 3 2400bit s 4 1200bit s 5 600bit s...

Страница 12: ...erred in a communication channel example In the serial port context 9600 baud means that the serial port is capable of transferring a maximum of 9600 bits per second TROUBLESHOOTING Error Reason Solut...

Страница 13: ...uick Disconnect DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes depending on the indicator model Figure 4 shows the pin assignment on the DB9 9 pin...

Страница 14: ...2 port and the indicator will respond with the current data Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overl...

Страница 15: ...IFICATIONS CONTACT US 2 5 2 ON OFF HOLD UNIT ZERO 8 4 5 Please e mail info selletonscales com for any sales related questions or call 844 735 5386 Don t forget to visit our website at www selletonscal...

Отзывы: