The desired value is also selected with the arrows. To set a
numeric value, it can be increased/decreased by single digits by
briefly pressing the respective arrow. Keep an arrow pressed, to
change the numeric value in steps of 10.
Press enter to confirm the entered value and leave the setting
mode. Press ESC if no value is to be changed and to leave the
setting mode.
To return to the display, press the dial symbol or wait for
30 seconds without making any input or touching the display.
User menu
Standard user parameters
are indicated in
bolt
and
underlined
.
Parameter
Designation
Value range
Description
Offset
Enabling offset
OFF
, ON
Enables/disables the
offset setting
Offset Val
Offset
-
100,0 …
0,0
(for unit %)
Sets the offset value
A reliable offset value
can only be set if the
current measured value
does not change or only
minimally changes over
a period of approx. 5
seconds..
Unit
Display toggling
%
, HRU
Sets the measuring
value to be displayed:
%
: Turbidity
HRU
: High Resolution
Unit
Independent from
display toggling, the
analogue output always
returns a signal
depending on the