MANUAL DE INSTRUCCIONES CODIGO 80222 REV A NOV-2007
(Sujetas a modificaciones sin previo aviso)
Pag.: 14
J
.P. SELECTA s.a.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
6.3 Control panel
1 .
Speed display (rpm)
2 .
Time display (min)
3 .
Increase speed value key
4 .
Decrease speed vaue key
5 .
Increase time value key
6 .
Decrease time value key
7.
Instant start/stop key
8 .
Start key
9 .
Stop key
6.4 Rear part
10. Power supply socket
11. Main switch
3
1
2
6
7
9
5
4
10
8
5 Introduction
The centrifuge Centro-4-BL is a tabletop centrifuge with a free maintenence
induction motor and electronic digital control of speed and working time,
controled acceleration and deceleralion, temporized mode from 1 minute to
60 minutes and safety electromecanic lock.
6 Equipment description
6.1 Equipment
The Centro-4- BL centrifuges have the following features:
-
Electronic digital control circuit board
-
Induction motor (brushless)
-
Swing out rotorto 4 tubes of 75 x 13 mm VAC
-
Electromagnetic safety lock
6.2 Features
-
Regulable speed untill 5000 rpm
-
Max RCF 2907 g
-
Speed reading on the display (resolution 10 rpm)
-
Working time reading on display (resolution 1 min)
-
Instant start/stop key to the set speed
-
Electronic lid opening
12
11