Operator
Passcodes and Operator Lock-Out
The instrument has two levels of passcodes to provide restrict Setup via the keypad.
Full Setup Passcode
(For calibration and Technician Access)
Setting a passcode for Full Setup restricts any access to Full Setup.
Full Setup passcode can also be used to access Safe Setup.
Safe Setup Passcode
(May allow limited Operator Access)
Setting a Passcode for Safe Setup restricts access to Safe Setup functions.
Setup Operator Lock-Out
If an attempt is made to enter Full or Safe Setup using an incorrect
P
asscode, the instrument will respond
with the message
ENTRY DENIED
and then the user will be returned to normal operating mode.
A
P
asscode counter has been set so that
only three failed attempts
can be made to access Full/Safe
Setup. On the fourth attempt the user will be 'locked out' of Full/Safe setup.
Should this occur the
ENTER PASS
prompt will not display, but instead the
ENTRY DENIED
message displays and returns the user to the normal operating mode. To rectify this issue the instrument
must be turned off. When the instrument is turned back on the
P
asscode counter is reset to zero
(allowing the user to enter the correct
P
asscode).
SAFE.PC
(Safe Passcode
for Operator access.
) allows partial access to Digital Setup (ie. only
non
-
calibration/trade critical settings can be changed).
For the Safe Passcode
(
SAFE.PC
)
to have any effect, the
FULL.PC
(Full
P
asscode
)
must also be set.
The default
P
asscode setting is 000000 that allows free access.
Any other number will restrict access.
Range:
000000
to
999999
Default:
000000
FULL.PC
(Full Passcode) can be set to restrict access to Full Digital Setup.
This
P
asscode is used to prevent unauthori
z
ed or accidental tampering in the instrument setup.
The default
P
asscode setting is 000000 allows free access.
Any other number will enable the
P
asscode functions and restrict access.
Range:
000000
to
999999
Default:
000000
It is important to note that when restricting Full access to Setup
the
P
asscode must not be
forgotten.
It is only possible to circumvent the
P
asscode
loss
at the factory
.
Care must be taken when setting the Full Digital Setup Passcode to ensure that the instrument does not
become permanently locked.
( Normally NOT USED )