7.3.3 Setting interval time for data storing:
(1) Press the
or
button to set minute or second. (see Fig.12)
(2)
Press the
/
button to increase / decrease value.
Note:
Set interval time range: 1
Second (00:01) to 60 minutes 59 seconds (60:59)
7.3.4 Set offset to compensate for probe errors:
The user can adjust the readings of thermometer to compensate for the errors against a specific
thermocouple.
(1) Press the
or
button to select thermocouple channel.
(see Fig.13)
Setting thermocouple channel blinks on the LCD screen.
(2) Press the
/
button to increase / decrease value.
Note:
The offset to compensate for probe errors range: ±5°C or ±9°F.
7.3.5 Set alarm point (only for channel T1):
Want to set window. (see Fig.14 or Fig.15)
(1) Press the
or
button to turn on or off.
(2) When setting on mode, press the
button
to set Hi and Lo limit. (see Fig.16)
Note:
When measuring thermocouple value over alarm point, the symbol
will blink
“
” or “
” on the LCD.
The beeper will make
“
beep-beep-beep
” sound.
The Lo alarm set cannot be greater than Hi alarm set.