User Manual – SEFRAM 53X2DC
33
Utility key 2/4
Enter self calibration
Self CAL
Go to previous menu
Previous Menu
Utility key 3/4
Select Hardcopy function
Function
Turn on/off inksaver
Ink Saver
Go to previous menu
Function
Save All
SaveImage/
SaveAll
Previous
Menu
On/ Off
Ink Saver
Off
Hardcopy
Previous Menu
Utility key 4/4
Select probe compensation signal
Wave Type
Set frequency for square wave
Frequency
→
VAR
Set duty cycle for square wave
Duty Cycle
→
VAR
Go to previous menu
Previous Menu
User Manual – SEFRAM 53X2DC
34
Default Settings
Here are the factory installed panel settings which
appear when pressing the Save/Recall key
→
Default Setup
.
Acquisition
Mode: Normal
Channel
Scale: 2V/Div
Invert: Off
Coupling: DC
Probe attenuation: x1
BW limit: Off
(GDS-1102, GDS-1062)
Channel 1 & 2: On
Cursor
Source: CH1
Cursor: Off
Display
Type: Vectors
Accumulate: Off
Grid:
Horizontal
Scale: 2.5us/Div
Mode: Main Timebase
Math
Type: + (Add)
Position: 0.00 Div
Measure
Item: Vpp, Vavg, Frequency, Duty Cycle, Rise
Time
Trigger
Type: Edge
Source: Channel1
Mode: Auto
Slope:
Coupling: DC
Rejection: Off
Noise Rejection: Off
Utility
Hardcopy: SaveImage,
InkSaver Off
ProbeComp: Square
wave, 1k, 50% duty cycle