User Manual – SEFRAM 53X2DC
71
5.
Press
When
repeatedly to
select the pulse condition.
Then use the Variable knob
to set the pulse width.
When <
20.0ns
VARIABLE
Condition
> , < , = ,
≠
Width
20ns ~ 200us
6.
Press
Slope/Coupling
to set
trigger slope and coupling.
Slope /
Coupling
7.
Press
Slope
repeatedly to
select the trigger slope,
which also appears at the
bottom of the display.
Slope
Range
Rising edge, falling edge
8.
Press
Coupling
repeatedly to
select the trigger coupling.
Coupling
AC
Range
DC, AC
9.
Press
Rejection
to select the
frequency rejection mode.
Rejection
Off
Range
LF, HF, Off
10.
Press
Noise Rej
to turn the
noise rejection on or off.
Noise Rej
Off
Range
On, Off
11.
Press
Previous
menu to go
back to the previous menu.
User Manual – SEFRAM 53X2DC
72
Manually triggering the signal
Note
This section describes how to manually trigger the
input signals when the oscilloscope does not
capture them. This section applies to the Normal
and Single trigger mode, since in the Auto trigger
mode, the oscilloscope keeps updating the input
signal regardless of the trigger conditions.
To acquire the
signal regardless
of trigger
conditions
To acquire the input signal
regardless of the trigger
condition, press the Force key.
The oscilloscope captures the
signals once.
FORCE
In the Single
trigger mode
Press the Single key to start
waiting for the trigger
condition. To break out of the
Single mode, press the
Run/Stop key. The trigger
mode changes to the Normal
mode.
SINGLE
Run/Stop