PrimeTest 350
Operating Instructions
- 53 -
Press
Class
(F2) to set the class of the appliance
under test. For a class II appliance the probe symbol
will be displayed at the top of the LCD display.
For both Class 1and Class II appliances fit the
Appliance mains plug into the EUT outlet. For Class 2
appliances only connect the Earth Bond lead to the
appliance. Press the green
START
key to begin the
test. The test will run for the duration set or until the red
STOP
key is pressed.
11.3 Substitute Leakage
Refer to Fig. 5
This test is applicable to Class I and Class
II equipment.
This test is used to verify that the mains supply pins
are adequately insulated from earth.
During this test, 40V A.C. is applied between the
earth pin and both the live and neutral pins of the
appliance mains supply plug.
Press
Setup
(F3) to set the Test Duration and
Pass/Fail limit. Use the arrow keys to highlight and set.
Press
OK
(F4) when complete.
Esc
(F2) will exit setup
mode without saving changes.
Press
Class
(F2) to set the class of the appliance
under test. For a class II appliance the probe symbol
will be displayed at the top of the LCD display.
For both Class 1and Class II appliances fit the
Appliance mains plug into the EUT outlet. For Class 2
appliances only connect the Earth Bond lead to the
appliance. Press the green
START
key to begin the
test. The test will run for the duration set or until the red
STOP
key is pressed.