PrimeTest 300
Operating Instructions
- 35 -
PATGuard software.
To insert a User Test in a new sequence follow the
step in section 9.1.4 to open the Inset a Test window
and highlight
User Test
. Press Select (F4) to insert a
User Test and display the Edit User Test screen shown
below.
The User Test Name will now be highlighted. For
convenience a selection of commonly used names
have been entered. Use the left/right arrows to scroll
through the factory-programmed names or enter a new
name using the keypad and edit function keys.
Now, using the down arrow, move the highlight bar to
the User Test Units field. Again, a selection of
commonly used units is available for convenience.
Select the required units using the left/right arrow or
enter new units using the keypad.
The screen below shows a User Test created to record
microwave leakage in mW/cm
2
.