Seaward PrimeTest 200 Скачать руководство пользователя страница 20

10

8.4

IEC Lead TEST, IEC

See Figure 6.

8.4.1

Select the IEC test function on the rotary switch  10 .

Do not connect the IEC lead to the PrimeTest 200, in preparation for
an IEC lead test, before the unit is switched on. Failure to comply
will result in an error message displayed on the LCD.

8.4.2

To test IEC leads connect the IEC socket side of the lead into the IEC
plug 14 . Connect the mains plug side of the lead into the EUT
socket  11 .

8.4.3

To start an IEC lead test press the        button  9 .

8.4.4

The earth continuity resistance for the lead is measured at ±200mA. 
The large display  2  indicates the highest digital and analogue earth 
resistance, 

. An earth continuity resistance >20 

will result in a FAIL.

An additional visual display of a 

or 

also indicates the PASS/FAIL

status.

Attention - 13A IEC or extension leads of different cable lengths will
have different resistances. As a rule of thumb:-

12 metres of 13A cable

0.2

6 metres of 13A cable

0.1

3 metres of 13A cable

0.05

8.4.5

The earth continuity resistance value is held on the LCD display for a 
period of 1 second.

8.4.6

An earth continuity resistance that FAILS will terminate the IEC lead test,
otherwise it will automatically sequence to the insulation resistance test.

During an insulation resistance measurement a high voltage is
applied between the LIVE/Neutral and common of the IEC lead.

8.4.7

The insulation resistance of the lead is measured over a period of 2
seconds. The large display  2  indicates digital and analogue insulation
resistance, M

, the small display  3  indicates a PASS/FAIL status. An

additional visual display of a 

or 

also indicates PASS/FAIL status.

8.4.8

Insulation resistance value is held on the LCD display for a period of 1
second. The displayed resistance value will flash if a FAIL and remain 
steady if a PASS. 

8.4.9

An insulation resistance that fails will terminate the lead test, a pass
will automatically sequence to the lead wiring test and the small digital 
display  3  will indicate “LEAD”.

8.4.10 The wiring of the IEC lead is tested over a period of 2 seconds. The 

large digital display  2  indicates a PASS/FAIL status for the lead. An 
additional visual indication of a 

or 

also indicates PASS/FAIL status.

The IEC lead wiring test checks for open circuit, short circuit or 
crossed wires within the IEC lead.

8.4.11 If this is the last IEC lead test or the only test then the results will remain 

displayed until another rotary switch or push button action.

8.5

Safety Test, Class II 

See Figure 7

8.5.1

Select the Class II     test function on the rotary switch  10  . The LCD
will display a probe symbol to indicate that a safety test lead is 
required. Ensure the IEC mains lead is disconnected from the
PrimeTest 200.

PrimeTest 200

Operating Instructions

Содержание PrimeTest 200

Страница 1: ...PrimeTest 200 Operating Instructions Bracken Hill South West Industrial Estate Peterlee Co Durham SR8 2SW England Tel 44 0 191 586 3511 Fax 44 0 191 586 0227 www seaward co uk sales seaward co uk cal...

Страница 2: ...Operating Instructions Primetest 200 Operating Instructions Primetest 200 Operating Instructions Primetest 200 Fig 1 Fig 2 Fig 3 Fig 4...

Страница 3: ...Operating Instructions PrimeTest 200 Fig 5 Fig 6...

Страница 4: ...Operating Instructions PrimeTest 200 Fig 7 Fig 8...

Страница 5: ...Operating Instructions PrimeTest 200 Fig 9 Fig 10...

Страница 6: ...Operating Instructions PrimeTest 200 Fig 11...

Страница 7: ...Operating Instructions PrimeTest 200 Fig 12B Fig 12A...

Страница 8: ...PrimeTest 200 Operating Instructions...

Страница 9: ...opying recording or in any manner without prior written consent from SEAWARD Electronic Limited This also applies to accompanying drawings and diagrams Due to a policy of continuous development SEAWAR...

Страница 10: ...PrimeTest 200 Operating Instructions...

Страница 11: ...otes These operating instructions are intended for the use of adequately trained personnel The PrimeTest 200 has been designed to make measurements in a dry environment It must not be used for making...

Страница 12: ...ring function and insulation resistance measurement function M Electric shock danger Where safe operation of the tester is no longer possible it should be immediately shutdown and secured to prevent a...

Страница 13: ...n up to 200 M resistance measurement 10 Measuring function selects the voltage up to 250 V AC DC measurement 11 Measuring function C selects the temperature up to 400 C measurement 4 2 LCD and Keymat...

Страница 14: ...Test 200 is switched on from power off position to another switch position an on load battery test occurs The result of this test determines the available battery capacity A healthy battery results in...

Страница 15: ...infinity Bargraph accuracy 10 Test voltage 500VDC at 1mA 750VDC for o c Test voltage accuracy 20 0 Test current 2mA for s c Input protection 0 5A 600 V quick blow fuse An input voltage of 30 V AC DC...

Страница 16: ...current measurement Measuring Range Resolution Accuracy Overload Protection 3 5mA 0 01mA 10 2 digits 5 40 N A 15 2 digits 0 5 Test voltage mains supply 230V 15 10 10A max Frequency 50Hz 1 Test duratio...

Страница 17: ...ck test lead to exposed metalwork on the EUT 8 1 6 To start a Class I safety test sequence press the button 9 8 1 7 The earth continuity resistance for the EUT is measured at 200mA The large display 2...

Страница 18: ...the test is inhibited 8 2 5 To start the differential leakage test press the button 9 Mains power is applied to the EUT for a period of 2 seconds 8 2 6 Where a longer differential leakage test is req...

Страница 19: ...o the PrimeTest 200 then PE and PN symbols will flash and the test is inhibited 8 3 5 The earth loop resistance is checked and an ohmic range selected on the large digital display This resistance valu...

Страница 20: ...nt a high voltage is applied between the LIVE Neutral and common of the IEC lead 8 4 7 The insulation resistance of the lead is measured over a period of 2 seconds The large display 2 indicates digita...

Страница 21: ...or pushbutton action 8 6 Touch Leakage Current See figure 8 Attention before a touch leakage current test is performed on a Class II EUT the operator must ensure that the EUT has already passed a Cla...

Страница 22: ...display 3 will revert to indicating a PASS FAIL status An additional visual display of a or also indicates the PASS FAIL status 8 6 10 The touch leakage current will remain displayed until another ro...

Страница 23: ...h potential Electric shock danger 8 8 1 Ensure the circuit or test object to be measured is voltage free Ensure the IEC mains lead is disconnected from the PrimeTest 200 8 8 2 Select the insulation re...

Страница 24: ...ure 12a and 12b 8 10 1 Select the voltage measurement function on the rotary switch 10 8 10 2 Plug the voltage measurement adaptor into the IEC mains input plug 14 on the end of the PrimeTest 200 8 10...

Страница 25: ...aning 9 3 1 Clean the external case of the PrimeTest 200 with a clean dry cloth 9 3 2 Avoid using solvents and abrasive scouring agents to clean the external case of the PrimeTest 200 9 3 3 Check the...

Страница 26: ...be displayed 9 6 Checking the EUT Socket Fuses 9 6 1 Plug the PrimeTest 200 into the mains power socket using the IEC lead provided 9 6 2 Select one of the three power tests differential leakage curre...

Страница 27: ...ery compartment and fasten in position with the battery cover captive screw 9 8 Calibration To maintain the specified accuracy of the measurement results the instrument must be recalibrated at regular...

Страница 28: ...rence Positions 90 E1 N A Supply Voltage At the Limits Specified E2 0 52 by the Manufacturer Temperature 0 C and 35 C E3 1 Absorbed Current of Maximum Absorbed Current Device on Test of the Device on...

Страница 29: ...90 E1 N A Supply Voltage At the Limits Specified E2 0 53 by the Manufacturer Temperature 0 C and 35 C E3 1 Absorbed Current of Maximum Absorbed Current Device on Test of the Device on Test E4 N A Low...

Страница 30: ...ositions 90 E1 N A Supply Voltage At the Limits Specified E2 0 56 by the Manufacturer Temperature 0 C and 35 C E3 0 81 Absorbed Current of Maximum Absorbed Current Device on Test of the Device on Test...

Страница 31: ...ons 90 E1 N A Supply Voltage At the Limits Specified E2 0 75 by the Manufacturer Temperature 0 C and 35 C E3 1 4 Absorbed Current of Maximum Absorbed Current Device on Test of the Device on Test E4 N...

Страница 32: ...h Earth Open Circuit On Flash Flash Neutral Open Circuit Err 1 The PrimeTest 200 has failed one of the Earth Relay Pre Check tests Contact your local service agent Err 2 The PrimeTest 200 has failed o...

Страница 33: ...PrimeTest 200 Operating Instructions...

Страница 34: ...PrimeTest 200 Operating Instructions...

Отзывы: