EX100/200 SERIES INSTRUCTIONS
Seametrics • 253.872.0284
Page 7
seametrics.com
“D”
It is important for accuracy that the sensor be inserted to
the correct depth into the pipe.
1. Please visit
seametrics.com
and select the
K-factor
Calculator
located at the bottom of the home page
to find dimension D (insertion depth setting)*.
2. Measuring from the outside of the pipe to the joint
in the housing, as shown in the diagram, adjust
the sensor to Dimension D and hand-tighten
compression nut.
3. Align the conduit housing with the centerline of the
pipe, as shown. Be sure the arrow on the housing
points in the direction of flow.
4. Check Dimension D one more time.
5. Fully tighten the compression nut.
Record your settings.
Once you have the meter set up and
operational, it is important to record you meter settings
and save them future reference.
K-Factor
___________________________________
Insertion Depth (Dim. D) __________________
* For pipe sizes larger than 50”, please consult factory.
strain
relief
FLOW
INSTALLATION
Caution!
Never attempt to remove a
flow sensor when there is pressure in the
pipe unless it is specifically designed for
hot tap installation and removal. Loosen
the compression nut slowly to release any trapped
pressure. If fluid sprays out when removing the
sensor, stop turning and depressurize the pipe. Failure
to do so could result in the sensor being thrown from
the pipe, resulting in damage or serious injury.
Depth Setting
Proper Depth Setting