DO2 INSTRUCTIONS
Seametrics • 253.872.0284
Page 12 seametricscom
MAINTANANCE
Maintenance
Sensor:
There are no user-serviceable parts. If problems develop with sensor
stability or accuracey, contact Seametrics. If the unit has been exposed to hazardous
materials, do not return without notification and authorization.
Cable:
Cable can be damaged by abrasion, sharp objects, twisting, crimping,
crushing, or pulling. Take care during installation and use to avoid cable damage.
End Connections:
The contact areas (pins & sockets) of the connectors will wear
out with extensive use. If your application requires repeated connections other
types of connectors can be provided. The connectors used by Seametrics are
not submersible, but are designed to be splash-resistant.
DO Window Cleaning:
During deployment the sensing window on the DO2 may
collect debris or growth. Heavy debris/growth can affect the accuracy of the DO
readings over time. Periodically check the window and clean if necessary. Note:
the DO window is very sensitive and prone to scratching, DO NOT use abrasive
cleaners. We recommend using a wash bottle and DI water to remove the loose
debris. If there is a stubborn growth over the window you may clean it with a
very soft cloth or Q-tip and a very light touch. Take care not to exert too much
pressure or scratch the window surface during cleaning. A mild detergent may
be used if necessary, such as Alconox. The main body of the unit may be cleaned
with a mild detergent as well.