background image

4.2. 

   

COUPLINGS

 

Vibration may cause failure if a quick change coupling is connected directly to the tool. To overcome this, connect a leader hose to  

 

the tool. A quick change coupling may then be used to connect the leader hose to the air line recoil hose. See fig.1 & 2. 

5.  OPERATION

 

‰

  

WARNING!

 Ensure you read, understand and apply safety instructions before use.

 

   Note:

 Numbers in brackets refer to item numbers in the parts diagram.

5.1.    

FLUX CHIPPER

5.1.1. 

  

Fit the spring (4) onto the rounded end of the chisel (15) before fitting the chisel into the barrel sleeve (12).

5.1.2.   

Fit the nut (20) over the sharpened end of the chisel (15), screw it onto the barrel sleeve (12) and tighten.

5.1.3.   

Connect the tool to the air hose.

5.1.4.   

To start the tool, depress the throttle lever. 

5.2.    

NEEDLE SCALER

5.2.1.   

To use as a needle scaler, unscrew the nut (20) and remove the chisel (15) and spring (4).

5.2.2.   

Fit the needle driver (14) and ensure the spring (18) is located in the needle tube (19).

5.2.3. 

  

Ensure the needle holder (17) is fitted with needles (16) such that the heads of the needles do not protrude above the surface of   

 

the needle holder.

5.2.4.   

Fit the needle holder (17) into the needle tube (19), screw the needle tube (19) onto the barrel sleeve (12) and tighten.

5.2.5.   

To start the tool, depress the throttle lever. 

 

8

  DO NOT

 allow the tool to free run for an extended period of time as this will shorten its life.

6.   MAINTENANCE

 

‰

  

WARNING!

 

Disconnect tool from air supply before changing accessories, servicing or performing maintenance.

 

 

Replace or repair damaged parts. 

Use genuine parts only. 

 

Unauthorised parts may be dangerous and will invalidate the warranty.

 

Note: 

 Numbers in brackets refer to item numbers in the parts diagram.

6.1. 

  

To replace the needles 

(16), unscrew the needle tube (19) from the barrel sleeve (12). Remove the needle holder (17) and replace  

 

the needles (16). (Use genuine Sealey needles only.)  Refit the needle holder (17) into the needle tube (19), screw it onto the barrel  

 

sleeve (12) and tighten.

6.2. 

  

Lubricate the tool daily with a few drops of Sealey air tool oil dripped into the air inlet.

6.3. 

  

Clean the tool after use. 

6.4.   

 

Loss of power or erratic action may be due to the following: 

 

Excessive drain on the air line. Moisture or restriction in the air pipe. Incorrect size or type of hose connectors. To remedy check    

 

the air supply and follow instructions in section 4. 

      

Grit or gum deposits in the tool may also reduce performance.  Flush the tool out with gum solvent oil or an equal mixture of 

 

SAE No 10 oil and paraffin. 

 

Allow to dry before use.  

6.5. 

  

For a full service contact your local Sealey service agent. 

6.6. 

  

When not in use, disconnect from air supply, clean tool and store in a safe, dry, childproof location. 

Original Language Version

© Jack Sealey Limited

   fig.1

   fig.2

SA52 Issue 3 (H,F)   24/01/20

Отзывы: