5.  Continue to press and hold the Verification Tester's 

WRIST STRAP test switch. Simultaneously, press 

and hold the Verification Tester's LOW BODY 

VOLTAGE test switch. The monitor's operator LED 

will remain green, and no audible alarm will sound. 

This verifies the operator circuit's low body voltage 

limit. 

Figure 11. Pressing the WRIST STRAP and LOW BODY 

VOLTAGE test switches simultaneously

6.  Continue to press and hold the Verification Tester's 

WRIST STRAP test switch. Simultaneously, press 

and hold the Verification Tester's HIGH BODY 

VOLTAGE test switch. The monitor's green operator 

LED will illuminate continuously, its red LED will 

blink. This verifies the operator circuit's high body 

voltage limit. 

Figure 12. Pressing the WRIST STRAP and HIGH BODY 

VOLTAGE test switches simultaneously

7.  Disconnect the mono cable from the monitor.

VERIFYING THE MAT CIRCUIT

8.  Connect the red test lead to red banana jack located 

at the top of the Verification Tester. 

9.  Disconnect the monitor's white mat monitor cord 

from its worksurface mat and turn it over to expose 

its 10 mm snap. 

10. Clip the red test lead's mini grabber to the 10 mm 

snap on the white mat monitor cord. 

11. Wait approximately 5 seconds for the monitor's mat 

LED to illuminate red and sound its audible alarm. 

12. Press and hold the Verification Tester's SOFT 

GROUND test switch. The monitor's mat LED will 

illuminate green, and its audible alarm will stop 

after approximately 3 seconds. This verifies the mat 

circuit's resistance limit. 

Figure 13. Connecting the Verification Tester to WS 

Aware Monitor's mat monitor cord

Monitor's

Mat Monitor Cord

Figure 14. Pressing the SOFT GROUND test switch

13. Disconnect the red test lead from the monitor's 

white mat monitor cord. 

14. Reinstall the white mat monitor cord to the 

worksurface mat.

Содержание CTE701

Страница 1: ...nds annual calibration of workstation monitors and the CTE701 Verification Tester The CTE701 Verification Tester meets ANSI ESD S20 20 and Compliance Verification ESD TR53 November 2022 The SCS CTE701 Verification Tester can be used with the following items Item Description 770067 WS Aware Monitor 770068 WS Aware Monitor CTC061 3 242 WW WS Aware Monitor CTC061 RT 242 WW WS Aware Monitor CTC062 RT ...

Страница 2: ...operator circuit when pressed F Soft Ground Test Switch Simulates a MAT PASS condition on the monitor when pressed G Wrist Strap Test Switch Simulates a OPERATOR PASS condition on the monitor when pressed H Test Limit DIP Switch Configures the test limits on the CTE701 Verification Tester I High Metal Ground Test Switch Simulates a TOOL FAIL condition on the monitor when pressed J High EMI Test Sw...

Страница 3: ...FF 14 ohms OFF ON OFF OFF 15 ohms ON OFF OFF OFF 16 ohms OFF OFF OFF OFF EMI The EMI high frequency signal is configured with switch 9 The CTE701 Verification Tester provides two different levels of high frequency signal elevated and normal Pressing the HIGH EMI pushbutton will load a high signal level within its range Pressing the LOW EMI pushbutton will load a low signal within its range Switch ...

Страница 4: ...OLTAGE test switches simultaneously 6 Continue to press and hold the Verification Tester s WRIST STRAP test switch Simultaneously press and hold the Verification Tester s HIGH BODY VOLTAGE test switch The monitor s green operator LED will illuminate continuously its red LED will blink and an audible alarm will sound This verifies the operator circuit s high body voltage limit Figure 6 Pressing the...

Страница 5: ...und 18 Set the variable DC power supply to 4 0 V The monitor s Iron LED should illuminate green and its audible alarm should stop 19 Disconnect the variable DC power supply from the monitor and ground Connect its positive terminal to ground and its negative terminal to the monitor s yellow alligator cord 20 Verify that the variable DC power supply is still set to 4 0 V The monitor s Iron LED shoul...

Страница 6: ...e 12 Pressing the WRIST STRAP and HIGH BODY VOLTAGE test switches simultaneously 7 Disconnect the mono cable from the monitor VERIFYING THE MAT CIRCUIT 8 Connect the red test lead to red banana jack located at the top of the Verification Tester 9 Disconnect the monitor s white mat monitor cord from its worksurface mat and turn it over to expose its 10 mm snap 10 Clip the red test lead s mini grabb...

Страница 7: ...ch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit 20 Press and hold the Verification Tester s METAL GROUND PASS test switch Simultaneously press and hold the Verification Tester s EMI LOW test switch The monitor s tool LED will remain green and no audible alarm will sound This verifies the operator circuit s low EMI volt...

Страница 8: ...s the tool circuit s impedance limit Figure 21 Pressing the METAL GROUND PASS test switch 5 Press and hold the Verification Tester s METAL GROUND FAIL test switch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit Figure 22 Pressing the METAL GROUND FAIL test switch 6 Press and hold the Verification Tester s METAL GROUND PAS...

Страница 9: ...necting the Verification Tester to the Ground Man Plus Workstation Monitor s operator jack Figure 26 Pressing the WRIST STRAP test switch Figure 27 Pressing the WRIST STRAP and LOW BODY VOLTAGE test switches simultaneously 4 Press and hold the Verification Tester s WRIST STRAP test switch The monitor s operator LED will illuminate green and its audible alarm will stop This verifies the operator ci...

Страница 10: ...GROUND PASS test switch The monitor s tool LED will illuminate green and its audible alarm will stop This verifies the tool circuit s impedance limit 12 Press and hold the Verification Tester s METAL GROUND FAIL test switch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit 13 Press and hold the Verification Tester s METAL G...

Страница 11: ... Battery LED illuminates red Open the compartment located on the back of the tester to replace the battery The tester uses one 9V alkaline battery Ensure that the battery s polarities are oriented correctly to avoid possible circuit damage Specifications Operating Temperature 50 to 95 F 10 to 35 C Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humid...

Отзывы: