SCS CTC337 Скачать руководство пользователя страница 4

TB-9055

 Page 4 of 6

© 2016 DESCO INDUSTRIES, INC.

Employee Owned

 

Specifications

Ground Impedance 

Alarm Level

1 to 20 ohms

Wrist Strap Alarm - 

Body Voltage

±2.5V on operator

Wrist Strap Alarm - 

Operator Loop 

Resistance

10 megohms on operator (plus 

two 1 megohm resistors in the 

end of the ground cord that 

attaches to the wristband)

Visual Alarm 

Indicator (LED)

Green: OK

Red: Fault Condition

Note: Operator LEDs red and 

green will not illuminate if wrist 

strap is not plugged into jack.

Audible Alarm

Buzzer fault condition if enabled

Operating 

Environmental 

Range

Temperature: 50 to 110ºF (10 to 

43ºC)

Humidity: 75% R.H. maximum

Dimensions

2.42" W x 2.45" L x 0.85" H

(61mm W x 62mm L x 22mm H)

Power Adapter

100-240VAC

50-60 Hz

Output: 12VDC @ 1.5A

Output Plug Polarization: 

Center Positive

Output Plug: 5.5 mm O.D. x 2.1 

mm I.D. x 9.5 mm L

Functional Operation

Ground Resistance Monitor

• 

Connect the ground wire to a known good electrical 

ground point.

• 

Connect the output wire to the tool if you intend to 

use it, otherwise leave it open.

• 

Connect the ground monitor input to the identified 

location.  Be sure to connect the monitoring input to 

different physical locations than the point used for its 

grounding. 

Person Voltage and Resistance Monitor 

Plug the dual conductor ground cord of the wrist strap 

into the jack on the front of the unit.  If the wrist strap 

is worn properly, a green LED will come on; otherwise 

red LED will be on.  In order to provide good ground 

connection, the wrist strap must be worn tightly.  The 

monitor’s green LED indicates a good connection.
If for whatever reason, while properly connected, the 

operator developed excessive body voltage (in excess 

of ± 2.5 Volts), the green LED may still be on, but the 

red LED will blink without alarm sound.  No LED will 

illuminate when the operator jack is left open.

Maintenance

Replacement and Removing Input Wrist Strap Dual 

Conductor Input Jack 

If the wrist strap input jack begins to show wear after 

heavy usage, the input jack can be replaced by ordering 

the SCS CTA252 Replacement Wrist Strap Jacks.  

Remove the input jack in the unit by removing the 

locking collar and two screws on front cover.  Slide the 

printed circuit board outside of the unit and unplug the 

pin connector with input jack board.  Plug in new input 

jack board assembly and carefully slide the board back 

into the housing.  Replace front cover and locking collar 

on the input jack.

Figure 4.  SCS CTA252 Replacement Wrist Strap Jacks

www.

.com

[email protected]

1.800.561.8187

Отзывы: