46
CHAPTER 5: MAINTENANCE
087-0038 REV L
SCOTT SAFETY
Figure 33
Sensor Pin Alignment
7
Replace back cover and install the seven (7) Torx screws. Hand-tighten the screws
taking care not to over torque as this can cause damage to the device housing.
8
Use the SGM software to set or verify settings for the newly installed sensor. See
See
“Configuring the Scott Gas Monitor Software” on page 23.
9
Perform Zero and Span Calibrations of the device in accordance with calibration
procedures. See
“Zero Calibration” on page 39.
Cleaning the Pump
This section describes how to clean the pump.
Warning:
If device does not operate in the manner described, do not use. Tag it out of
service. Failure to ensure it is properly operating may result in serious injury or
death.
Required Equipment:
•
10ml Syringe
•
Tygon Tubing
•
T8 Torx head screw driver
•
Isopropyl Alcohol (IPA)
•
Can of compressed air
2 Alignment
2 Recepticle
Содержание Protege
Страница 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 2: ......
Страница 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Страница 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Страница 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Страница 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Страница 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Страница 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Страница 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Страница 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Страница 103: ......
Страница 104: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...