Calibrating the Device
39
SCOTT SAFETY
087-0038 REV L
Zero Calibration
This section describes how to perform Zero Calibration.
1
Verify ambient air is free of toxic or combustible gases. If air is not free or cannot
be verified free, obtain a Zero Air gas cylinder.
2
Power on the device.
3
Press and hold the Right Operation Button until PLEASE WAIT displays. It is in
progress. See
.
Figure 25
Zero Calibration Display - In Progress
Warning:
To prevent an erroneous Zero Calibration from occurring, the device
displays IS AIR CLEAR? If the device is indicating a gas reading that is 50% or
higher of an alarm set point for any of the gases it is equipped to detect. You have 20
seconds to acknowledge the alert. If the alert is not acknowledged within 20 seconds,
the device powers off. Do not continue with Zero Calibration until air has been
verified free of toxic and combustible gases. Failure to properly perform a zero
calibration may result in injury or death.
Baseline readings for a Zero Calibration are 0 for H
2
S, CO, and LEL, and 20.9% for
O
2
.
Table 11
Recommended Calibration and Test Matrix
ITEM
FREQUENCY
DETAILS
Zero Calibration
•
Daily
•
When baseline readings are incorrect or
suspect
•
Prior to a Span Calibration
Span Calibration
•
CAL EXPIRED is displayed
•
After installing new sensors
•
When a Bump Test fails
Bump Test
•
Daily
•
When a sensor is exposed to a high
concentration of gas
•
When submerged in water
•
When exposed to mechanical shock, such as
being dropped
Содержание Protege
Страница 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 2: ......
Страница 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Страница 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Страница 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Страница 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Страница 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Страница 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Страница 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Страница 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Страница 103: ......
Страница 104: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...