Calibrating the Device
43
SCOTT SAFETY
087-0038 REV L
Bump Testing
This section describes how to perform Bump Test.
Equipment required:
•
Calibration gas – Verify concentration level exceeds set points of monitor and
expiration date of cylinder has not passed.
•
Tygon tubing – 2 feet of 3/16" ID
•
Regulator – Set to provide flow at 0.5LPM
•
Calibration Adapter – Included with the device
1
Verify the concentration level of the target gas in the cylinder exceeds alarm
settings of the Protégé as set using the SGM software. See
Gas Monitor Software” on page 23.
If needed.
2
Attach the Regulator to the gas cylinder and verify cylinder pressure.
3
Connect Tygon tubing to both the Regulator and Calibration Adapter.
4
If installed, remove optional pump from device and restart the device.
5
Attach Calibration Adapter to device and apply gas. See
Figure 30
Bump Test - Calibration Adapter Attached
6
Verify device responds to target gas and activates the visual, audible, and vibrating
alarms.
7
Verify readings on the LCD /-20% of actual target gas concentration.
8
Turn off gas cylinder and remove Calibration Adapter.
Warning:
If the device fails to activate all alarms in a timely manner, the device must
be serviced prior to usage. If the device fails to read target gas concentration within
+/-20%, perform a Span Calibration. Failure to do so could lead to death or injury.
Содержание Protege
Страница 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 2: ......
Страница 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Страница 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Страница 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Страница 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Страница 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Страница 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Страница 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Страница 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Страница 103: ......
Страница 104: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...