Operation & Maintenance Manual
Manual #087-0007G, 10/2003
Troubleshooting
General Difficulties
Problem
Solution
Transmitter output is
Transmitter was probably zero
continuously negative.
calibrated with hydrocarbons
present. Apply a source of clean
air to the sensor (p/n 077-0019)
and zero instrument.
Response to gas is sluggish.
Check dust filters and replace, if
necessary (see "
Maintenance:
Replacing Dust Filters
").
Display is blank, although
Adjust display contrast (see
transmitter is producing
"General Operation: Adjusting
an output signal.
LCD Display Contrast"
).
Alarm set point cannot be
Lower alarm reset point, then try
lowered.
lowering set point.
"MAINT" Indicator displayed
The MAINT indicator appears when the
reference detector (IR1) signal drops
below 1.5 volts, peak-to-peak. This will
occur when enough dirt/dust
contamination has collected in the
sensor body to reduce the amount of
reflection below 30% of its original value.
If this occurs, the mirror assembly will
likely require cleaning. See "
Periodic
Maintenance: Cleaning the Mirror
Assembly
".
High Zero Reading
Follow the periodic maintenance
procedures detailed at the beginning of
POWER
SWITCH
Place in the OFF position
before removing Display /
CPU board stack. Ensure
power switch is in
the OFF position prior to
replacing board stack.
Note: DO NOT
change out
CPU or power
supply boards.
These boards
are factory
calibrated as
a matched set.