46
Model 4688-IR / 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
Displayed
As:
Description
Min
Max
Default
Value
MEM
Region
MODBUS
Register
Address
Data
Type
[BITS]
Assigned
Parameter
#
-10
%LEL(or
equiv.
%V/V)
0 %LEL
-5 %LEL
CO2
-2.0% v/v
0.0 %v/v
-2.0% v/v
AF.OP
Fault Alarm Options.
This parameter
should never be set to "0" on the Model
4688I-R.
0= non-latching,non-failsafe
1=non-latching, failsafe
The failsafe option keeps the relay energized
in the non-alarm state and de-energizes it in
alarm (and power failure). The latching
option requires you to manually reset the
alarm indicator and relatch after an alarm,
once concentration has fallen to (or below)
the alarm's setpoint value.
0
1
1
PMEM
40273
Unsigned
Byte
[8]
22
AL.IP
Alarm Inhibit Period
- Specifies the number
of minutes which the alarm inhibit period will
remain on before automatically timing out
and re-enabling alarm relays.
9 min.
120 min.
9 min.
PMEM
40271
Unsigned
Integer
[16]
23
AL.tO
Alarm Test Over-Ride
- Used to test each
alarm indicator and relay individually, or in
groups (regardless of the alarm inhibit state).
Activates alarm indicators and energizes non
failsafe relays (de-energizes failsafe relays).
Note that this test has no affect if the alarm
is currently active and cannot be used to turn
off an alarm. The state is reset to 0 at startup
and is defined below:
0 - None 4 - Fault
1 - AL1 5 - AL1, Fault
2 - AL2 6 - AL2, Fault
3 - AL1, AL2 7 - AL1, AL2, Fault
NOTE: To prevent false alarms at the
receiver, set the receiver channel to alarm
inhibit prior to making these adjustments.
0
7
0
VMEM
0
1
0 (%LEL)
CO2
1
1
1(% v/v)
USER MENU - Alarm Relay Parameters
(continued…)
Unsigned
Byte
[8]
25
Sensor Concentration Units
- Specifies the
units in which to display the measured
concentration, full scale value, alarm set
points, alarm reset points, and negative
concentration limit:
0=%LEL
1=%V/V.
The transmitter will automatically change the
display will automatically change the
displayed decimal precision based on this
setting. When %LEL is chosen, the decimal
precision is automatically changed to 0 (no
decimal point). When %v/v is chosen, the
decimal precision is changed to 1 (0.1). The
decimal precision precision may be
overridden using the SR.dP parameter.
SR.CU
PMEM
40336
USER MENU - Sensor Parameters
AF.NL
Maximum Negative Drift Fault Alarm
- At
or below this concentration the unit will
display the fault indicator, open the fault
relay, and output the programmed milliamp
fault level. Units are the same as the
displayed reading in RUN mode.
PMEM
40331
REAL
[32]
21