INCU-C Series V2018.1.1
17
8 .
Setting of the Other Functions
8.1 Setting of over temperature alarm value. (Alarm Deviation Temperature)
Press the Set key and
enter password “5”. Press the Set key again to enter into the over
temperature alarm value. Press the Increase or Decrease key to change the required
value of temperature deviation. Press the Set key again to confirm and store the new
settings.
Simplified instructions for over temperature alarm settings (ADT)
When the temperature exceeds this over temperature alarm set point, the alarm sounds
continuously and the LED alarm indicator on the control panel lights.
Alarm can be muted by pressing the Temperature button.
8.2 Real time clock setting
Press the SET key and enter t
he password “4”.Press the SET key again to enter into
time set menu. Press Increase or Decrease key to set the right time of day.
Press the SET key again, the display shows the year, press the Increase or Decrease
button to change it to the actual year. Press the SET key once more, the display shows
the month, press the Increase or Decrease button to change it to the actual month,
press the SET key one more time to the day of the month, press the Increase or De-
crease button to the right date.
Press SET key to store and return to normal display
Simplified instructions for Clock settings.
Date
Month
Attention:
1. When
this ADT value is set on “0”, the alarm function is deactivated.
2. Changing the set (PV) temperature value, may immediately activate the
alarm if as a result, the adjustment makes the difference exceed the de-
viation limit.
3. When the actual temperature is back within limit(s), the unit automati-
cally stops the temperature alarms.
4
Time
Year
Date set and store
5
Over temperature
Alarm value