Maintenance
A7-10
ENVI-MAG Manual
Data does not dump
RS-232 cable
not connected
Connect cable.
Computer not
communicating
1. Run the computer
communications
program.
2. Check that you are
using the correct
serial port on your
computer.
Baud rate incompatible
Set baud rates on the
computer and ENV-
MAG console to be
identical.
No such data in
instrument
1. Make sure that you
are asking for data
that you have mea-
sured, i.e. asking for
base or grad when t-
fld recorded.
2. Select: dump all
records.
No matching line
number
Set proper line number
or disable line by line
dump.
Base-station correction
does not work
Base unit and mobile
unit not properly
interconnected.
Interconnect units via
Null Modem.
No matching date or
time
Correction must be
made externally on a
computer.
Cannot recall data
No respective line
or station (or time,
if base-station).
Set recall parameters
to match survey
parameters,
i.e. LN, SEP, STN, SEP
may be out a digit.
Problems
Possible Causes
Possible Solutions
Содержание ENVI-MAG
Страница 1: ...Part Number 788 711 November 1994 ...
Страница 12: ...Preface x ENVI MAG Manual ...
Страница 34: ...Instrument A3 2 ENVI MAG Manual Figure A 9 The ENVI MAG console 2 1 3 4 7 8 9 5 10 6 3 11 ...
Страница 76: ...Instrument A3 44 ENVI MAG Manual ...
Страница 108: ...Setup A4 32 ENVI MAG Manual ...
Страница 160: ...Reference A8 8 ENVI MAG Manual ...
Страница 166: ...Applications B1 4 ENVI MAG Manual Figure B 4 The intensity of the Earth s magnetic field ...
Страница 184: ...Applications B2 8 ENVI MAG Manual ...
Страница 199: ...Field example ENVI MAG Manual B3 15 Example Figure B 28 Columbia test site total field contoured data ...
Страница 200: ...Applications B3 16 ENVI MAG Manual Figure B 29 Columbia test site vertical gradient contoured data ...
Страница 202: ...Applications B3 18 ENVI MAG Manual ...
Страница 252: ...Index ENVI MAG Index ...
Страница 253: ......
Страница 265: ...Index Index xiv ENVI MAG Manual ...