USER MANUAL
4
NOTE:
This centrifuge has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the centrifuge is operated in a commercial environment. The centrifuge generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the user
manual, may cause harmful interference to radio communications. Operation of centrifuge in a residential
area is likely to cause harmful interference in which case the user will be required to correct the
interference.
3. Required Operational Condition
3.1 Basic operational Conditions
(
1
)
Power:
•
Single-phase, 220V-240V, 50Hz/60Hz, 8A
;
•
Single-phase, 110V-120V, 50Hz/60Hz, 12A
;
(
2
)
Ambient temperature: 2
℃~
40
℃
.
(
3
)
Relative humidity:
≤
80%.
(
4
)
No vibration and airflow around.
(
5
)
No electric dust, explosive and corrosive gases around.
3.2 Transport and storage condition
(
1
)
Storage temperature: -
40
℃~
55
℃
.
(
2
)
Relative humidity:
≤
93%.
4. Installation
This section describes the instructions that you should abide when install the centrifuge to ensure your
safety and the optimum performance. Before moving the centrifuge, the rotor must be removed.
WARNING
:
Improper power supply may damage centrifuge.
Make sure the power source conforms to the required power supply before connecting.
Содержание SCI636
Страница 3: ...USER MANUAL 2...
Страница 4: ......