USER MANUAL
Rotor type
ID
code
Tube/bottle
Adapter
Max.
speed
(rpm)
Max.centrifugal radius
r
max
(cm)
Max. RCF
Rcf (×g)
1
AS30-2
2/1.5ml
×
30
6000
10
4020
0.2ml
×
30
A02P2
6000
8.5
3415
0.5ml
×
30
A05P2
6000
9
3618
2
AS60-2
2/1.5ml
×
60
6000
10
4020
0.2ml
×
60
A02P2
6000
8.5
3415
0.5ml
×
60
A05P2
6000
9
3618
3
AS18-5V
5
mlV
×
18
6000
10
4020
4
AS12-PCR8
8-PCR
×
12
6000
10
4020
0.2ml
×
96
6000
10
4020
5
AS6-50V
50mlV
×
6
6000
10.7
4306
6
A30-15
15mlV
×
30
4500
R1=14.2
R2=12.2
3210
2760
7
A8-50
15mlV
×
16
50mlV
×
8
5000
12.4
3460
8
SE4-100A
100ml
×
4
4000
15.9
2840
85ml
×
4
A85P100
4000
15.9
2840
50mlV
×
4
A50VP100
4000
15.9
2840
15mlV
×
8
A15VP100
4000
15.1
2700
3
~
10ml
×
8
A10P100
4000
14.7
2630
9
S2-MP
(dimension
)
mm
MTP
(128
×
85.6
×
15)
4000
12.1
2160
Cell culture
(128
×
85.6
×
21)
4000
12.1
2160
DWP
(128
×
85.6
×
45)
4000
12.1
2160
Kits
(128
×
85.6
×
60)
4000
12.1
2160
* :
15mlV means 15ml conical tube, as well as 5mlV and 50mlV.
Table 11.1 Rotors and adapters
11.1.3 Notice
(1) The centrifuge rotor can separate sample which density lower than 2.0g/ml, if the samples density is
over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
(
2
)
To prevent corrosion, remove the rotor from rotor chamber if do not use for a long time, then detach the
rotor lid and upside the rotor down to dry the tube holes.