stuck between the rotor and the shaft. Double check and clean it if necessary. 

 

Rotate the rotor slightly with your fingers to check if the rotor vibrates. If so reinstall the rotor again. 

 

Rotate the nut clockwise using the wrench to tighten the rotor to the shaft firmly. 

 

Close the rotor lid, firmly tighten clockwise the lid to the rotor and ensure is in position. Close the door 

and then start running. 

 

The method of removing the rotor is as same as the above mentioned by turning the locking nut 

counterclockwise. 

 

8.1.2 

Set the operation parameters 

Press  the 

  key  to  select  required  parameters.  The  parameter  can  be  modified  when  the  parameter  is 

flashing.  Clockwise  rotate  the  parameter  key

to  increase  parameter  value;  counter-clockwise  rotate  the 

parameter  key 

  to  decrease  parameter  value.  Parameter  key

rotate  faster,  parameter  value  increase 

faster. The minimum speed increment is 100 rpm, the minimum time increment is 1 second. 

1

Set the speed

 

 

Press the key

  until the speed rpm is displayed. 

 

When the speed is selected, the speed symbol will flash the speed value. 

 

The minimum speed value you can set 300rpm, the minimum increment is 100rpm. 

 

Rotate  parameter  key 

clockwise  to  increase  speed  value,  rotate  the  parameter  key 

anti-clockwise to decrease speed value. 

 

You can speed-up set the speed value by rotating parameter key

  quickly. 

 

There  is  a  circulating  function  to  increase/decrease  the  speed  values.  Rotate  the  parameter  key 

clockwise 

  to  change  settings  from  small  →  large  →  maximum  →  minimum.  Rotate  the 

parameter key anti-clockwise

  to change settings from large → small → minimum → maximum. 

2

Set the time 

 

Press key

, time value flashes in the time setting mode. 

 

Rotate the parameter key

  to set running time from 10 seconds to 99 minutes. 

 

When time displays HD, this is a continuous running mode. 

(3)    Set acceleration and deceleration 

 

Press key

, acceleration value flash, press the 

  again, the value will increasing.the value will 

change from 1 to 9, then from 9 to1. 

1 acceleration: the slowest acceleration; 9 acceleration: the fastest acceleration. 

 

 

Press key

, deceleration value flash, press the 

  again, the value will increasing.the value will 

change from 0 to 9, then from 9 to 0. 

0 deceleration: free braking;     

1deceleration: the slowest deceleration; 9deceleration: the fastest deceleration. 

 

Содержание SCI636

Страница 1: ...SCI636 Multi purpose Centrifuge Before using centrifuge please carefully read this user manual for its efficient operation and safety...

Страница 2: ...ect the rotor 8 7 5 Symmetrically load centrifuge tubes in rotor 8 8 Operation 8 8 1 Normal Operation 8 8 2 RCF Operation 12 8 3 Pulse Operation 12 9 Maintenance 12 9 1 Cleaning 12 9 2 Consumables 13...

Страница 3: ...USER MANUAL 2...

Страница 4: ......

Страница 5: ...y be reproduced or transmitted without prior written permission of original manufacture We can not be responsible to inform at real time if the outline and specifications of centrifuge is subject to c...

Страница 6: ...death CAUTION Possible damage to instrument Caution notes indicate any condition or practice which if not strictly observed or remedied could result in damage or destruction of the instrument NOTE Not...

Страница 7: ...orizontal before running Make sure the angle between the door and cover is greater than 70 degrees when open the door Be careful not put your fingers or hands between the door and cover when the door...

Страница 8: ...emperature detector Imbalance protection Automatic internal diagnosis Power requirements Single phase 220V 240V 50Hz 60Hz 8A 110V 120V 50Hz 60Hz 12A Dimensions mm L 280 W 364 H 266 Weight 30kg Additio...

Страница 9: ...ich case the user will be required to correct the interference 3 Required Operational Condition 3 1 Basic operational Conditions 1 Power Single phase 220V 240V 50Hz 60Hz 8A Single phase 110V 120V 50Hz...

Страница 10: ...at least 30cm behind it to guarantee the cooling efficiency 4 Keep away from heat or water to avoid sample temperature issues or centrifuge failures 4 2 Connection of the power cord and grounding WAR...

Страница 11: ...Open lock key Press the key to open the door The key is not available when the centrifuge is running 3 Start Stop key Press the key to start running The centrifuge will brake to stop running if press...

Страница 12: ...is 9 memory program is 7 When speed symbol is rotating indicating the centrifuge is running the rotation is faster the speed is higher Temperature only displays the temperature of chamber and can not...

Страница 13: ...erwise the rotor may be moved off while rotating and cause damage of the centrifuge and rotor Firmly tighten the rotor door with rotor 8 Operation CAUTION Do not push or lean against the centrifuge wh...

Страница 14: ...00rpm Running time 30 minutes The door lock is released 8 1 1 Load and replace the rotor Figure 8 2 Load the rotor CAUTION Attach the rotor to the rotor shaft Ensure the rotor is in position and conne...

Страница 15: ...peed symbol will flash the speed value The minimum speed value you can set 300rpm the minimum increment is 100rpm Rotate parameter key clockwise to increase speed value rotate the parameter key anti c...

Страница 16: ...normal operation mode and run according to the new value If the set time value has been modified the operation time is not affected and will continue 3 Warning display If an error occurs during the o...

Страница 17: ...the power switch and load the rotor to the shaft tighten the rotor lid and make sure it is in secured position and then close the door 2 The centrifuge gets into preparation mode and displays last ru...

Страница 18: ...mple leaks in the rotor rinse the rotor with water Apply a thin coat of silicon grease to the rotor when it is completely dry The rotor should be regular maintenance recommend to cleaning it each 3 mo...

Страница 19: ...n the table firmly 2 Check if the centrifuge grounded properly Use multi meter to check if is short circuit between the power cord grounding pin and the motor shaft If yes indicating grounded properly...

Страница 20: ...llowable imbalance is exceeded some wrong in the drive system Balance the sample checked by professional person E 10 86 Read the service manual Contact with service center Table 10 1 Possible problems...

Страница 21: ...eed the allowable maximum speed of rotor tube and adapters etc be care that the allowable maximum speed of some adapters are lower than the rotor s maximum speed 11 1 The rotor instructions 11 1 1 Rot...

Страница 22: ...50mlV 4 A50VP100 4000 15 9 2840 15mlV 8 A15VP100 4000 15 1 2700 3 10ml 8 A10P100 4000 14 7 2630 9 S2 MP dimension mm MTP 128 85 6 15 4000 12 1 2160 Cell culture 128 85 6 21 4000 12 1 2160 DWP 128 85 6...

Страница 23: ...5 Bio safe seal ring The rotor is sealed by bio safe structures achieved using three high temperature rubber seal rings The seal rings may fall off or aging after several autoclaving need to be repla...

Страница 24: ...ove screw nuts and inner covers to prevent from deformation or rupture 3 Wait until the sterilizing chamber cools down to the room temperature before the tubes are removed 11 2 4 Condition and life ex...

Страница 25: ...Rotor kits with cover fixing clips and O ring 6000rpm PCR8 12 used with D1536 D1536R 19400002 AS6 50V Rotor kits with cover fixing clips and O ring 6000rpm 50mlV 6 used with D1536 D1536R 19400002 A30...

Страница 26: ...the following conditions even if within the guarantee period expires 1 Failures caused by incorrect installation 2 Failures caused by rough or improper handling 3 Failures caused by conveyance or rel...

Страница 27: ...USER MANUAL 22 SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 United States Office 1 860 436 9221 Fax 1 860 436 9745 Sales service contact info scilogex com www scilogex com...

Отзывы: