![Scientific SM6015A Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/scientific/sm6015a/sm6015a_user-manual_1217324042.webp)
Scientific
SM6015A
42/44
400nF
40.00nF - 399.99nF
0.25%+2 digits
±0.0025
Series
40nF
4.000nF - 39.999nF
0.25%+2 digits
±0.0025
----
4nF
400.0pF - 3.9999nF
0.25%+2 digits
±0.0025
Parallel
400pF
40.00pF - 399.99pF
0.35%+3 digits
±0.0035
Parallel
40pF
0.00pF - 39.99pF
1.25%+5 digits
------
Parallel
Impedance
(
(
(
(
Z
)
)
)
)
and Phase Angle
(
(
(
(
θ
)
)
)
)
Range
Display Range
Accuracy
Recommended
Equivalent
Mode
Ze
θ
e
1
0
0
H
z
/1
0
k
H
z
10M
Ω
4.000M
Ω
- 10.000M
Ω
3.00% +5 digits
±1.75°
Parallel
4M
Ω
400.0k
Ω
– 3.9999M
Ω
1.25% +3 digits
±0.75°
Parallel
400k
Ω
40.00k
Ω
– 399.99k
Ω
0.35% +2 digits
±0.25°
Parallel
40k
Ω
4.000k
Ω
– 39.999k
Ω
0.25% +2 digits
±0.15°
Parallel
4k
Ω
400.0
Ω
– 3.9999k
Ω
0.25% +2 digits
±0.15°
-----
400
Ω
40.00
Ω
– 399.99
Ω
0.25% +2 digits
±0.15°
Series
40
Ω
4.000
Ω
– 39.999
Ω
0.35% +2 digits
±0.25°
Series
4
Ω
0.4000
Ω
– 3.9999
Ω
1.00% +5 digits
±0.60°
Series
0.4
Ω
0.0000
Ω
– 0.3999
Ω
3.00% +5 digits
---
Series
*Note: Accuracy of De is assessed when De <0.5
Equivalent Series Resistance
Accuracy of equivalent series resistance is calculated according to the below formula:
Содержание SM6015A
Страница 1: ...Hand Held LCR Meter SM6015A User Manual ...