Generally, Series mode is better for components with low impedance (blow 100
Ω
),
while parallel mode for components with high impedance (over 10k
Ω
). For the
components with the impedance between the two limits, equivalent mode has little
effect on the testing result.
Selecting Measurement Mode
Long press P<->S button, “PAL” on LCD indicates parallel equivalent mode and
“SER” means series equivalent mode.
Default Equivalent Mode
In this mode, default equivalent mode varies with primary parameter:
For capacitors and resistors, default equivalent mode is PAL; for inductors, SER.
6.11 Utility Menu (UTIL)
The LCR meter has a built- in utility menu that allows you to configure some user
preferences and settings. The buttons used to set and control the menu are colored
in blue. There are three buttons: UTIL ,
▲
,
▼
. User can configure the beep tone,
auto power-off
timing, store/restore power-on state, view the battery voltage, etc.
Entering Utility Menu
Long press UTIL button will enter utility menu. Primary display is menu option and
secondary display is the current settings or parameters configured for the selected
option. After the entrance into utility menu, the default menu item display will show
“bEEP”.
Configuration and Settings
The following contents are included in the utility menu:
Table 2-Utility Menu Options and Settings
MenuOptions
Settings/Parameters
bEEP
ON / OFF
AoFF
5 / 15 / 30 / 60 / OFF
PuP
PrE / Set
dEF
yES / NO
bAtt
Battery Voltage
Uses of these menu options are as follows:
Control beep sound: ( bEEP: beep sound)
Set auto power-off: (AoFF: auto power off)
Set power-up state: (PuP: power-up state)
Reset to default settings: (dEF: default settings)
Indicate battery voltage: ( bAtt: battery voltage).
By default, press the button UTIL to change or select a different menu option. To
change the settings or parameters, use the
▲
and
▼
arrow keys. For each UTIL
button press, the meter will traverse through each menu options and will repeat itself
in the following order: