Maintenance
There are no user serviceable part inside
.Your
30 V, 5 A Power
Supply is thoughtfully engineered for ease of use, accuracy and reliability. The
instrument is carefully tested and calibrated using standards traceable to National
Laboratories.
Take care of your instrument by cleaning the exterior of the instrument regularly
with a dusting brush. Dirt which is difficult to remove on the casing & plastic parts,
can be removed with a moist cloth ( 99% water, 1% mild detergent) spirit or washing
benzene (petroleum ether) can be used to remove greasy dirt. The display may be
cleaned with water or washing benzene (but not with spirit-alcohol solvents), it must
then be wiped with a dry clean lint-free cloth. Under no circumstances the cleaning
fluid should get into the instrument. The use of cleaning agents can attack the
plastic & paint surfaces.
The power line fuse is located on rear panel on lower right side. In case, the
instrument does not show any sign of working, no LED is lit or there is no display
immediately switch OFF the mains power switch of the instrument and unplug the
mains cord from the mains socket. With the help of small flat blade screwdriver
remove the fuse cap of the fuse holder, located just below the socket. There is one
spare fuse kept in the fuse cap, replace it for the defective one. Press the cap so
that it locks in place . The rating of the fuse is 3 A , 250 V, sloblow , 5x20 mm glass
fuse. Do not use a fuse with a higher value
other wise it may damage the
instrument in case, the mains voltage goes much higher than the rating of the
mains fluctuation of 10%.
PSD3005
PSD3005
Power Line Fuse Replacement
±
11
User Manual PSD3005