KDS-1037
-
4
-
Specifications
CO
2
range
0 – 100,000 ppm (0 – 10%)
Accuracy
(at standard 1 atm):
0 ppm to 10,000ppm : 1,000ppm or 10% of value, whichever is
greater
10,000ppm to 50,000ppm : 20% of value
Over 50,000ppm : qualitative only
Output voltage
Output impedance
0 – 4 V
1kOhm
Resolution using 12 bit AD
30.5 ppm CO
2
Calibration information
Slope: 25,000 ppm/V, offset 0V
(0V=0 ppm, 4V=100,000 ppm)
Warm up time
10 minutes (maximum)
(faster when air currents are provided)
Response time
90% of full-scale reading in 60 s
Normal operating temperature
range
25
5°C
Operating hum. range
5 - 95 % (non-condensing)
Revision : Dec. 2014
© 2014 KoreaDigital Co. Ltd.
#804 Ace Twin Tower 2, Guro, Seoul, 152-779 South Korea
• TEL (+82) 2-2109-8880 • FAX (+82) 2-2109-8881
Direct : (+82)2-2109-8839
www.sciencecube.com
This product is to be used for educational purposes only. It is not appropriate for industrial, medical,
research, or commercial applications.
Our products and the contents are subject to change without any notice.
In consequence we cannot assume responsibility for any consequential or other damage resulting from the
use of this instrument.
ScienceCube are registered trademarks or trademarks of Korea Digital Co., Ltd.
Other product and brand names may be trademarks or registered trademarks of their respective
organizations.
Note :
This product is to be used for educational purposes only. ScienceCube CO2 Gas
Sensors are not designed nor recommended for any industrial, medical, or commercial process
such as life support patient diagnosis, control of a manufacturing process, or industrial testing of
any kind.