11. Specifications

12. Serial Number Plate

Applicable Standards

EN ISO 15883 parts 1 & 2

Height (Full height unit)

850 mm

Height (Short height unit)

830 mm

Width

600 mm

Depth (Door closed)

600 mm

Depth (Door open)

1200 mm

Weight

90 kg

Required clearance, top, sides and rear

>10 mm

Maximum Running Noise

< 78 dB(A)

Inlet water connections

G 3/4”

Inlet water pressure

2-5 bar

Drain

3/4”

Maximum water flow to drain

47 l/min

Maximum water discharge temperature to drain

95°C

Maximum water hardness

30.3dH, 31.6 US GPG, 540 PPM 

Maximum water conductivity

844 µS/cm

PH range

>6.8 and < 8.5

Water softener salt capacity

1.0 kg

Equipment installation category

II

Voltage

230 – 240 vAC +/- 10%

Frequency

50 Hz or 60 Hz, * see serial number label for 
requirements specific to your unit.

Rated load

2.5 kW

Circuit breaker

13 amps

Operating temperature range

+5°C to +40°C

Maximum relative humidity

80% for temp up to 31°C
50% for temp up to 40°C

Maximum operating altitude

2000 m

Equipment pollution degree

2

Maximum deviation from plane horizontal surface

2 mm

Detergent requirements:

Single solution – Scican HIP chemical only. 
Not supplied with machine.

Detergent consumption per cycle

108 ml

Consumption and processing temperatures

wash: 1.2% @ 70°C
rinse: 0.15% @ 60-94°C

When ordering supplies, spare parts or requesting service, please ensure that the 
information contained on the serial number plate is available (Model number, serial 
number etc.).

The serial number plate is located at the bottom left on the rear panel of the HYDRIM M2 
G4 unit. A small label is located on the left hand side of the chemical door.

Содержание HYDRIM M2 G4

Страница 1: ...HYDRIM M2 G4 Operator s Manual Mode d emploi Bedienerhandbuch Manual del operador Manuale d uso HYDRIM M2 G4 Operator s Manual 95 113643 Rev 1 0 Copyright 2013 SciCan Ltd All rights reserved...

Страница 2: ...e HYDRIM M2 G4 35 7 5 Filter and Wash Arm Maintenance 36 7 6 Annual Maintenance 36 8 Unit Testing and validation 37 9 Troubleshooting 39 10 Warranty 41 11 Specifications 42 12 Serial Number Plate 42 T...

Страница 3: ...nd replacement instructions must be followed for the product to perform as designed All trademarks referred to in this manual are the property of their respective owners Contents of this manual are su...

Страница 4: ...rface and or Hot Steam Caution Risk of electrical shock Disconnect power supply Caution Potential hazard to the operator The following symbols appear in this book This situation may lead to a mechanic...

Страница 5: ...es and mouth Never sit stand or lean on the open door The unit may tip forward causing injury Always turn the unit OFF before adding softener salt adding solutions or performing routine maintenance to...

Страница 6: ...rtaken before starting the machine See section 6 for care and maintenance details A separate pre installation checklist should have been supplied by your dealer Please review this prior to approving i...

Страница 7: ...ethylene engineered foam and wood pallets Some municipalities operate packaging collection points with dedicated disposal bins to accommodate various materials What to do with your old unit A decommis...

Страница 8: ...M M2 G4 such as the chamber show any of these spots or stains a water test may be required to determine the cause The installation of a water treatment system may be required to reduce the amount of d...

Страница 9: ...top maximum 1 kg 2 2 lbs with salt 8 Close by screwing the lid tightly back into place and run a complete cycle with no instruments An improper seal can lead to corrosion Please note The water test s...

Страница 10: ...ps scissors curettes and other stainless steel instruments free of deep cavities For best results instruments processed in the HYDRIM M2 G4 should have the following properties Heat resistance to a te...

Страница 11: ...PN 01 110409S Closed hinged instruments cannot be cleaned Place hollow lumens suction tubes and other hollow instruments in the rack for vertical instruments PN 01 01 110411S Do not place hollow lumen...

Страница 12: ...sinfection process may be achieved by means of the A0 method which uses knowledge of the lethality of the particular process at different temperatures to assess the overall lethality of the cycle and...

Страница 13: ...commended loading of general instruments and the worst case loading for which the unit has been type tested and validated Handle all instruments with care to prevent personal injuries through puncture...

Страница 14: ...pluggers curettes excavators etc For example Bone files pluggers curettes excavators etc For example dental handpieces turbines hollow lumens etc Handpiece holder washes or thermally disinfects the ou...

Страница 15: ...allation technician may be required and is recommended If one of these factory validated loads is used by the dental office and this can be documented by pictures and or appropriate lists an initial o...

Страница 16: ...ts Quarter size 1 4 Part number 01 109967S Rack for four 11 x 14 280 x 368 mm cassettes for upper or lower trolley Full size 1 1 Part number 01 109965S Statim 5000 basket Part number 01 107241 Hinged...

Страница 17: ...HYDRIM M2 G4 Touchscreen overview Settings Unlock door Network connectivity green when active USB status green when active Programs Water softener detergent level and door status 5 Instructions for Us...

Страница 18: ...creen Instructions Instructions Delay Remote Access Prepare for Shipping Set Printer Language Country Date Time Unit No Network Setup Reset Drying Counter Reset LCS Counter Screen Saver Temperature C...

Страница 19: ...tup menu 5 9 Setting the time 1 Date Time 2 Scroll to Date Time and select TIME 3 From the TIME screen use the keypad to set the time Press EN to save and to return to the Setup menu NOTE If the HYDRI...

Страница 20: ...ng the keypad select a maximum of 3 digits to be used as the unit s identifier number Press EN to save and to return to the Setup menu 5 12 Resetting the drying counter The drying counter must be rese...

Страница 21: ...these steps NOTE Turning OFF the button sound does NOT turn off other alarms and cycle notification beeps 1 2 Scroll to Beep ON OFF and select 3 Use to scroll through your ON or OFF options and select...

Страница 22: ...centers Should you need to adjust the contrast for your office follow these steps 1 2 Scroll to LCD Contrast and select 3 Use to scroll through your contrast options When you have found the contrast...

Страница 23: ...screen you can assign up to four PINs To assign a PIN follow these steps 1 2 Scroll to User and select 3 To assign a user PIN select User PIN and use the numeric keypad to enter a number up to 4 digi...

Страница 24: ...located at the back of the unit To connect your HYDRIM to a network using a router follow these steps 1 Connect your network cable to the Ethernet port at the back of the unit If your office uses a ro...

Страница 25: ...Connection time will vary depending on your network speed and making an initial connection can take longer 5 24 Connecting to a wireless network The HYDRIM can be configured for wireless use by connec...

Страница 26: ...jected cycle completion 8 The cycle is complete When the cycle is complete the program name and Cycle Complete is displayed and the unit beeps three times Press OK to open the door Use gloves to unloa...

Страница 27: ...menu functions To access the user or setup menus follow these steps Chamber cleaning This cycle is used to periodically remove hardwater deposits from the chamber walls and racks Pour 1 0 litres of vi...

Страница 28: ...ot Instruments that have not completed the cycle should not be used and should be reprocessed When replacing the kickplate ensure that the grounding wire is attached Grounding wire Ring 5 Instructions...

Страница 29: ...2 The unit will record up to the last five successful cycles and the last five incomplete cycles If you select a cycle from the list it will display cycle information in a format similar to how it wou...

Страница 30: ...rom the web portal 3 When the activity light on your USB storage device stops blinking or the USB icon on the LCD turns from a flashing green to a solid grey remove the USB storage device and transfer...

Страница 31: ...l Printer if connecting a printer or USB Flash MSD if connecting the SciCan Data Logger Press the back arrow to return to the Setup Menu 4 In the Setup Menu select Baud rate 5 Select the rate required...

Страница 32: ...undamaged gloves made of Natural or butyl rubber nitrile or neoprene Surgical gloves are ideal Always wear eye protection Protect exposed skin on arms If solution is spilled on clothing remove the ite...

Страница 33: ...cleaning solution follow these steps 7 2 1 Replacing the HIP Cleaning Solution The HYDRIM M2 G4 uses a 3 8L 1 US gallon container of HIP Cleaning Solution Part number CS HIPL Only use the recommended...

Страница 34: ...tion container by rotating forwards into the correct location in the solution door see picture 6 11 Close the cleaning solution door 12 Press OK on the touch screen The machine display will indicate P...

Страница 35: ...log the unit The water softener salt is also available from SciCan Please contact your local dealer 7 3 Changing the Air Filter When the message Replace Air Filter appears the HEPA filter is in need o...

Страница 36: ...ce call was performed Filter Maintenance Inspect the filter in the bottom of the chamber daily for debris and clean if necessary To clean remove the filter rinse under a tap and replace Ensure that th...

Страница 37: ...icians Failure to maintain this equipment may invalidate the results of the following testing regime The recommended periodic testing protocols are as follows Commissioning Installation testing and va...

Страница 38: ...eck for debris See section 6 5 Door seal check Visually check door seal for debris or wear If seal is worn or damaged call your dealer for technical service Load carrier check Visually check load carr...

Страница 39: ...ater is not hot enough Check that the incoming hot water temperature is in the recommended range Check that the hot and cold hoses are not reversed No detergent message comes on when there is still ch...

Страница 40: ...vice CF13 Temperature validation error Second temperature sensor readings are out of limits Call for service CF14 Water temperature too high Water in prewash stage too hot Check water connections at b...

Страница 41: ...invoice identifies the item by serial number and clearly states the date of purchase No other validation is acceptable After one year all SciCan s warranties and other duties with respect to the quali...

Страница 42: ...0 Hz see serial number label for requirements specific to your unit Rated load 2 5 kW Circuit breaker 13 amps Operating temperature range 5 C to 40 C Maximum relative humidity 80 for temp up to 31 C 5...

Отзывы: