![SciCan HYDRIM L110w Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/scican/hydrim-l110w/hydrim-l110w_service-manual_1217173027.webp)
Hydrim L110w Service Guide 95-109977-1
…
27
Use a Water Test Kit part #01-108305S to determine if salt is required.
Use the up and down arrows to select the desired salt regeneration setting.
Press
default
to return to the factory default.
Press the back arrow to return to the Technician Menu.
Select cycle for dosing setting you wish to
change.
Use the up and down arrows to select the desired dosing setting for the cycle
selected. (Do Not change default setting unless directed by SciCan)
Press
default
to return to factory default.
Press the back arrow to return to the Technician Menu.
Prewash dosing Min Value 0 – Max Valve 15s, Default Value 5s
Wash dosing Min Value 0 – Max Value 30s, Default Value 14s
HD Wash dosing Min Value 0 – Max Valve 30s Default 15s
Final rinse dosing Min Value 0 – Max Value 15s, Default Value 10s
Set Regeneration
0
Default 0
Chemical Setup
Chemical Setup
Prewash
Rinse
Set HIP Dosing
Prewash
5s
Default 5s
Wash
High Temp. Wash