SD-438 EU EN R5
5
Manganese
< 0.01
mg/L
> 0.04
Chloride
< 20
mg/L
> 30
*For water, milligram per liter (mg/L) equals parts per million (ppm)
If full water analysis results in parameters exceeding problem level, then a water treatment system
is highly recommended. The HYDR
IM
washer-disinfectors are fitted with an additional water inlet,
specifically for external water treatment systems (such as deionised or RO).
There may be instances where parameters are below the stated problem levels, however staining
still occurs. A water treatment system may be needed to prevent the staining – see SD-447 Water
Quality Document available on mySciCan for additional information.
The HYDR
IM
is also equipped with an air gap / anti-suction device to prevent backflow of dirty
water into the water supply.
•
Water hardness
o
Hot water
o
Cold water
•
Water conductivity
•
pH value
o
Action required (if any):
Internet Connection
!
/
"
Install a Cat 5 or 6 Ethernet port
Note 1: The port should be installed in a cabinet adjacent to the unit.
Note 2: Ensure that the port has access to the office network and the internet.
Note 3: Although the unit can be connected to a bridge for wireless access, it is recommended to use hard
wired connection to minimize risks with network connections.
Note 4: If the office has purchased (or plans to purchase) other internet connectivity (G4) products, install
multiple Ethernet ports to accommodate other G4 products from SciCan.
HIP/HIP Ultra cleaning solution
!
/
"
Order Cleaning solution
HYDR
IM
C61wd G4: HIP Ultra cleaning solution (CS-HIPC-U)
HYDR
IM
M2 G4: HIP cleaning solution (CS-HIPL) or HIP Enzymatic (CS-HIPL-E)
Note 1: HIP, HIP Ultra or HIP Enzymatic cleaning solution is required to run the machine and is not included
with the machine. Please ensure cleaning solution has been provided prior to installation.
Compressed Air (HYDR
IM
C61wd G4 LCS Model ONLY)
!
/
"
Availability of a DN 7.2 connection for compressed air
Air pressure must be between 1 bar (15 PSI) - 10 bar (145 PSI)
Distance between unit installation site and compressed air connection must be within 2 m