SD-438 EU EN R5
2
Installation Configuration
The washer-disinfector may be installed under-bench or free standing dependent on space and
service availability. Please select appropriate model number for the installation configuration.
The HYDR
IM
C61wd G4 configuration options are:
NOTE
: LCS Models must also have a compressed air supply within 2 m of installation site
Option 1: Cabinet/ Sterilization Center
Option 2: Sliding Shelf Option 3: Counter Top
The HYDR
IM
M2 G4 configuration options are:
Option 1: Cabinet/ Sterilization Center (M2 G4 Short model)
Option 2: Free Standing (M2 G4 Standard model)
Space Requirements
!
/
"
•
Ensure adequate space is available to install the unit
Note 1: Although not required, if the HYDR
IM
is installed under a counter, it is recommended to allow at least
10 mm space at the top and both sides of the unit. This will facilitate installation levelling, and service access
to the HYDR
IM
.
Note 2: Avoid installing the HYDR
IM
in direct sunlight.
Note 3:
Important!
Ventilation during drying is via the front of the machine. Some increase in humidity may
be apparent during this drying phase and it is imperative that sufficient air circulation is available surrounding
the machine to enable this humidity to be dissipated. Failure to provide adequate ventilation may lead to
equipment or cabinetry damage (depending on installation configuration).
Note 4: The drying system has a filtered air intake at the front of the unit. The free movement of air to this
Power connection
Hot Water
connection
Cold Water
connection
Drain
connection
598 mm + 10mm
820
m
m
598 mm + 20mm
850
m
m
598mm
598mm
520
m
m
+
10
m
m
598mm + 20mm
526mm
520mm
+
10
m
m
598mm + 20mm
526mm
520
mm
598mm
526 mm