SciCan HYDRIM G4 Скачать руководство пользователя страница 1

SD-438 EU EN R5 

 

 

 

 SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada. M3B 3P9 

 

 

HYDR

IM

 G4 Washer-Disinfector  

Pre-Installation Instructions 

 

To  ensure  the  optimum  performance  of  SciCan’s  washer-disinfector  models  it  is  imperative  that  the 
correct  services  are  made  available  and  that  the  positioning  of  the  unit  does  not  compromise  the 
function of the unit in any way. 
 
Prior to installation it is important that the site be inspected and all the appropriate conditions are met 
to ensure the safe and efficient operation of the washer-disinfector. 

 
Practice details and summary results. 

 

Practice/Office location details

 

Customer Name: 

 

 
 
 

Address: 

 

Tel: 

 

Signature: 

 

 

Installer/Pre-installation inspector

 

Name: 

 

 
 
 

Dealer details: 

 

Tel: 

 

 

Conditions approved/rejected for installation subject to checklist below (

!

/

"

)

 

 

Name: 

 

Signature: 

 

Date: 

 

On behalf of (Dealer): 

 

Содержание HYDRIM G4

Страница 1: ...oes not compromise the function of the unit in any way Prior to installation it is important that the site be inspected and all the appropriate conditions are met to ensure the safe and efficient oper...

Страница 2: ...ounter it is recommended to allow at least 10 mm space at the top and both sides of the unit This will facilitate installation levelling and service access to the HYDRIM Note 2 Avoid installing the HY...

Страница 3: ...cold water feed valve so that both machine pipes can be connected to the cold supply The machine will not work with only one supply attached as it requires pressure in both supplies to activate the f...

Страница 4: ...o Sealed edges o Heat resistance o Sound structural integrity o Post installation access o Flatness See note 3 above Water Quality The quality of the water being used in the HYDRIM to clean the medica...

Страница 5: ...ort Note 1 The port should be installed in a cabinet adjacent to the unit Note 2 Ensure that the port has access to the office network and the internet Note 3 Although the unit can be connected to a b...

Страница 6: ...ls D13 12A 200 230 V 50 Hz 1 50 m 5 ft 2 7 kW D01 12A 230 240 V 60 Hz 1 50 m 5 ft 2 5 kW D03 13A 230 240 V 60 Hz 1 50 m 5 ft 2 5 kW D02 13A 230 240 V 50 Hz 1 50 m 5 ft 2 5 kW D04 13A 230 240 V 50 Hz 1...

Страница 7: ...feeds are installed Drain pipes are prepared Electrical requirements are met Cabinetry floors can withstand operation of the unit Water quality is acceptable Internet connection installed HIP HIP Ult...

Страница 8: ...for drying Total water consumption per cycle with drying P1 P2 11 4 L 1 5 L P3 30 3 L 2 5 L Maximum heat transmitted through fascia 1200 W Water softener salt capacity 0 5 kg 1 1 lbs Equipment instal...

Страница 9: ...nge 6 8 and 8 5 Water volume per process stage 8 L 1 L P3 has additional 8 1L for drying Total water consumption per cycle with drying P1 16 L 2 L P2 32 L 4 L P3 40 L 5 L Maximum heat transmitted thro...