ESIRIS -
- Version 1.0 Date: 22.09.00
Laser error:
In case of laser problems the sytem will read out the error codes of the laser. The
following lines gives you a short explanation of the meaning of each errorcode. If the
service need more information, please take a look at the Optex manual from Lambda
Physiks for detailed information.
Errorcode 1:
Check error occurred (statical error)
Errorcode 2:
Time out error (operation error: laser device is to be switched off)
Errorcode 3:
Laser head chamber open or temperature to high ( more then 60 °C)
Errorcode 4:
External interlock circuit open
Errorcode 5:
Electronics chamber open or temperature to high ( more then 60°C)
Errorcode 6:
Reservoir pressure out of range ( less than 2.9 bar or more then 4.5 bar)
Errorcode 7:
Pressure is not reached in the present time limit (time out error)
Errorcode 8:
Heating circuit error (less then 5.8V or more than 7.5V)
Errorcode 9:
HV charging time exceeded (more than 4.5 ms)