- Version 1.0 Date: 22.09.00
3. Installation requirements and preparations
Installation of
SCHWIND ESIRIS
is performed through trained service personnel. To
enable a proper installation, some prerequisites have to be fulfilled, which are
described below:
Working Room
The room, where
ESIRIS
will be installed, should have a minimum of 3 x 4 m (12
m²). At the outer side of the entrance door, a
laser warning lamp
and the
official
laser warning symbol
have to be installed. The warning lamp has to lighten, during
the laser is in use. A switch, which interrupts the laser, if the door is opened, can be
connected through the interlock input at the device. But this is only recommended
when it is guaranteed, that the door is not opened during treatment.
Inside the laser room explosive or inflammable material must not be stored. Near the
opening door, protection glasses according to DIN EN 207 have to be available in
sufficient quantity. (Protection glasses have to be available in following
specifications: D L9 / IL2 calculated by DIN EN 207, according to Laser Vision).
Nitrogen gas bottles, necessary for use of the laser can be stored inside the laser
room. They must be fixed, to avoid falling down.
Important note:
Do not use any liquid containing ammoniac to clean the room. Ammoniac
vapours attenuate the laser energy and create varying ablation rates and
treatment results!