Operating Manual - SPORÉCUP XTR2
Selecting an area of stimulation
Depending on the selected program you can choose between upper and lower
limbs.
Press the
key to change the pulse width of the selected programs. The pulse
width is increased for lower limbs.
Programs with a possibility of upper and lower limbs stimulation:
P1, P5, P6, P7, P9, P10, P15-P29
Recommendation:
Higher pulse width for lower limb stimulation
Lower pulse width for upper limb stimulation
Setting the intensity
Start the stimulation by increasing the intensity with the key of the channel(s)
connected to the electrodes.
key to increase the intensity
key to decrease the intensity
The intensity of each channel is displayed on the
screen. The intensity can be set between 0 and
100 mA. By keeping the intensity keys pressed, the
intensity will increase/decrease more rapidly.
Terminating the stimulation
You can terminate the stimulation at any time by pressing the
key. When the
therapy time is completed the stimulation terminates automatically. If you press
the key the stimulation stops and the XTR2 switches off.
Locking the key pad
Choose the desired program by pressing the
key (see chapter selecting a
program).
Press and hold for 3 seconds the
key and the right key to lock the stimulator
on the chosen program. The XTR2 can now only be used with the chosen program.
A key symbol appears on the display adjacent to the program number. All keys
except the intensity and the ON/OFF keys are locked into this mode. Use the same
key combination to unlock the stimulator.