13
English
General information
Adjustments, modifications and repairs
The manufacturer is only responsible for the safety and performance of the XTR2
when readjustments, alterations and repairs are carried out by authorised persons
and when the XTR2 is used in accordance with the operating instructions.
Warranty
Legal right of warranty is applied according to German Civil Code.
Guarantee
The manufacturer issues a guarantee of 12 months from the date of purchase.
The guarantee does not apply in the following cases:
»
damage due to improper handling
»
defects the customer is aware of on the date of purchase
»
damage caused by the customer
»
or wearing parts and consumable supplies like, for instance, battery, cables and
electrodes.
Maintenance and cleaning
No special cleaning or care agents are required for the XTR2. Clean the stimulator
with a soft, fluff-free cloth. Please ensure that no moisture permeates the
stimulator. If moisture does permeate the stimulator, a technical check must be
carried out before re-use.
Classification
In accordance with the Law on Medical Devices, the XTR2 is classified as being a
Class IIa medical device.
Safety checks (§11 MPBetreibV)
[Medical Devices Operator Ordinance]
As an operator, you are legally obligated to perform safety checks for this product
every 24 months (in accordance with generally accepted technical standards).
The safety checks comprise:
1. Checking the accompanying documents to ensure that the instructions for
use and the medical device log are present
2. Checking the equipment to ensure it is complete
3. Visual inspection
-
for mechanical damage
-
of all lines and plug connectors for damage