Pointoselect Digital DT   |   Instruction Manual 

 

 

Manufacturer 

 

Environment 
Do not dispose of the appliance with the normal household waste at the end of its life, 
but hand it in at an official collection point for recycling. By doing this you will help to 
protect the environment. 

 

4. 

Technical Specifications 

 
Acupuncture point finder with automatic and manual location functions and integrated electrical 
stimulator. 
 

Power supply:  

 

9 V battery 

Dimensions:    

 

approx. 115 x 60 x 30 mm 

Weight:  

 

 

approx. 123 g without accessories 

 

Point finder 

Current consumption:  

approx. 17 mA (without signal tone) 

  

approx. 30 mA (with signal tone) 

 

Stimulator 

Current consumption:  

approx. 28 mA (under 20 k

 

load at 128 Hz and 120 μs)

 

Output voltage:  

 

200 V (under 20 k

 load) 

Frequency range:  

 

2-128 Hz 

Pulse width:    

 

60-

120 μs

 

 

 

Impulse form:  

 

 
Maximum stimulation pulse under 20 k

 load 

 

 

 
 

Electric current according to resistive load 

 

 
 
 
 
 
 

Содержание Pointoselect Digital DT

Страница 1: ...Pointoselect Digital DT Digital Point Finder for Auricular and Body Acupuncture With Integrated PuTENS Stimulation Item no 200506 Instruction Manual Item no 100858 ...

Страница 2: ...surement 6 Automatic Point Detection 6 Manual Point Detection 7 Gold silver Point Detection 7 Stimulation 7 Adjusting the volume of the acoustic signal 8 Switching OFF the device 8 Changing the battery 9 Short Instruction to find an ear point 9 7 General Information 9 Adjustments Modifications and Repairs 9 Warranty 9 Guarantee 9 Maintenance and Cleaning 10 Classification 10 Information on Battery...

Страница 3: ...r or on children Always store the Pointoselect Digital DT in its case to protect it from damage and dust Warning The Pointoselect digital DT should not be used close to other electronic devices e g mobile phones in order to avoid interference If this is not possible the device must be observed during operation to ensure that the respective functions are operating correctly The application of the P...

Страница 4: ...matic and manual location functions and integrated electrical stimulator Power supply 9 V battery Dimensions approx 115 x 60 x 30 mm Weight approx 123 g without accessories Point finder Current consumption approx 17 mA without signal tone approx 30 mA with signal tone Stimulator Current consumption approx 28 mA under 20 kΩ load at 128 Hz and 120 μs Output voltage 200 V under 20 kΩ load Frequency r...

Страница 5: ...the Pointoselect digital DT Take particular care of the connecting pins of the hand probe while plugging the hand probe in and out Note that the hand probe can only be connected to its socket in one direction Follow the arrow on the hand probe and on the socket of the device The patient needs to hold the hand electrode and the acupuncturist needs to touch the patient Starting the device Start the ...

Страница 6: ...the hand probe are required for measurements Body Exclusively the inner sensor is used for measurements in the body mode Gold and silver point detection is not active and therefore not displayed Measurement The lower the sensitivity is few bars on the display and deep acoustic signal the more relevant the point is for treatment Many bars on the display and high acoustic signal symbolize high sensi...

Страница 7: ...cted the mode Go only gold points will be detected and displayed If you have selected the mode Si only silver points will be detected and displayed Please Note Silver and Gold points can only be detected if you have selected an mode Stimulation Entering the stimulation mode The stimulation settings can be modified in the S mode Press the M key to access the stimulation mode S Following parameters ...

Страница 8: ...d outer sensors The parameters and the intensity used are the ones set up previously in the S mode The letter S for stimulation the intensity in V and the frequency in Hz are displayed on the screen The red LED of the hand probe is lighting up The intensity can be modified during stimulation pressing the left keys on the device or the keys of the hand probe The frequency can be modified during sti...

Страница 9: ... probe at a 90 degree angle to where you want to find a point on the ear Put the probe near a point and apply a little pressure If you can reduce the sensitivity display and sound at least to half of the original value then you are exactly on the point If you choose a manual mode select with M key you have to reduce the sensitivity slowly with the minus key on the probe until you have reached at l...

Страница 10: ...n points you re contributing to protecting our environment and the earth s natural resources Batteries are made from important resources and chemicals including lead cadmium zinc lithium and mercury If batteries are disposed of as normal municipal household waste they ll be taken to a landfill site and those resources will be lost and will contribute to the pollution of the environment Recycling i...

Страница 11: ... with the accessories listed in the chapter Delivery Content of this manual The manufacturer cannot guarantee the security of the device if not used with original accessories 8 Delivery Content Item no Description Quantity 200506 Pointoselect digital DT 1 200507 Hand probe 1 250200 Hand electrode 1 602000 Battery 1 106242 5 24 cable 1 106752 Belt clip 1 100858 Instruction manual 1 ...

Страница 12: ...struction Manual Manual version 09 2014 Distributor schwa medico GmbH Wetzlarerstr 41 43 35630 Ehringshausen Deutschland www schwa medico com export schwa medico de Pierenkemper GmbH Am Geiersberg 6 35630 Ehringshausen Deutschland ...