Calibration of the NSP-S 140 by the customer
Connect the measuring instrument to the electrical interface and
read the voltage value from the various clamping positions.
Clamping positions:
•
Open module --> Apply opening pressure 12 -25 bar and read
the voltage value on the measuring device.
•
Clamped module without clamping ring and without turbo U
A
--
> depressurized, clamped via spring. Read the voltage value.
•
Clamped module with clamping ring (pallet) and without turbo
--> spring-loaded clamping ring inserted. Read the voltage
value.
•
Clamped module with clamping ring and with 25 bar turbo U
B
--
> clamping ring inserted and actuated with 25 bar hydraulic
pressure (turbo). Read the voltage value.
The voltage values are used to determine the clamping position of
the module. The tolerance values must be added or subtracted to
these voltage values.
Example:
U
A
--> clamped without clamping ring: 2.096 V+T
A
(0.32 V) = 2.416
V
U
B
--> clamped with clamping ring and 25 bar turbo 2.898 V
–
T
B
(0.27 V) = 2.628 V
The exact limit can be assumed to be 2.6 V.
The voltage value for the clamping position "Open module" is
between 9 ..... 10 V.