Operation
ProLab 2000
82
ba75577e03
11/2009
4.9.2
Automatic storage at intervals
The storage interval (
Interval
) determines the chronological interval
between automatic storage processes. Each storage process transmits
the current dataset to the interface at the same time.
Configuring the
automatic storage
function
Settings
You can configure the automatic storage function with the following
settings:
1
Open the menu for automatic storage with
<STO
_
>.
SAMPLE ID
1
Interval
30 s
Duration
180 min
Continue
00d03h00min
00d15h
34min
<STO_>
Menu item
Possible
setting
Description
SAMPLE ID
1 ... 999
Determine the sample
designation for the series
of datasets.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min,
30 min, 60 min
Storage interval.
The lower limit of the
storage interval can be
restricted by the number of
free storage locations.
The upper limit is restricted
by the storage duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time
the automatic storage
should be terminated.
The lower limit of the
storage duration is
restricted by the storage
interval.
The upper limit is restricted
by the number of free
storage locations.